![JGR MBR5 User Manual Download Page 71](http://html.mh-extra.com/html/jgr/mbr5/mbr5_user-manual_647213071.webp)
MBR5 User Manual
MBR5 Rev 007
65
SWL?
Query returns the current wavelength number in
“x.x” format, valid returns are 1.3 for 1310nm, 1.4 for
1480nm, 1.5 for 1550nm, 1.6 for 1625nm and 0.8 for
850nm.
Interface:
GPIB and RS-‐232/USB
Syntax:
“SWL?”<CR>
Returned Format:
<x.x><NL> where <x.x> = “0.8”, “1.3”, “1.4”, “1.5”, or
“1.6”
SSC
The SSC command switches the meter to the next
wavelength or to a specific wavelength using the
position number/sequence as reference. The
position of a laser in a particular meter will depend
on the number of lasers in the meter.
Interface:
GPIB and RS-‐232/USB
Syntax:
“SSC”<CR> (switches to next wavelength)
“SSC x”<CR> (where x = 1, 2, 3, or 4). Example, “SSC
2” switches to the second wavelength (ie on a 2-‐
wavelength MBR5 1310/1550, "SSC 2" would switch
to 1550nm, however on a 1310/1480/1550/1625
"SSC 2" would switch to 1480nm)
Returned Format:
None
SSC?
Query returns the current wavelength position
number in “x” format.
Interface:
GPIB and RS-‐232/USB
Syntax:
“SSC”<CR> (switches to next wavelength)
“SSC?”<CR>
Returned Format:
<x ><NL> where < x> = “1”, “2”, “3”, or “4”