![JGR MBR5 User Manual Download Page 67](http://html.mh-extra.com/html/jgr/mbr5/mbr5_user-manual_647213067.webp)
MBR5 User Manual
MBR5 Rev 007
61
[:POWER]:REFerence:AWL
Syntax
[:POWER]:REFerence:AWL
Function Takes a relative power reference at current detector with
current channel, for all wavelengths.
Example
REF:AWL
takes the reference for all wavelengths.
[:POWER]:REFerence:ALL
Syntax
[:POWER]:REFerence:ALL
Function Takes a relative power reference at current detector for all
wavelengths and all channels.
Example
REF:ALL
takes the reference for all wavelengths and all channels
[:POWER]:REFerence:SAVe
Syntax
[:POWER]:REFerence:SAVe
Function Saves Insertion loss, BR0 and Setup via loss reference values to
flash. These values can be restored on the power on if the meter
is factory set for this function.
Example
REF:SAV
saves all reference values to flash.
[:POWER]:REFerence:REStore
Syntax
[:POWER]:REFerence:REStore
Function Restores Insertion loss, BR0 and Setup via loss reference values
from flash.
Example
REF:RES
reads and restores IL, BR0 and Setup via loss values.
[:POWER]:REFerence:CLEar
Syntax
[:POWER]:REFerence:CLEar
Function Restores default Insertion loss, BR0 and Setup via loss reference
values on the meter.
Example
REF:CLE
restores default reference values.