MBR5 User Manual
MBR5 Rev 007
28
Table 6: User Menu Options
Options
Note
REMOTE MENU
JGR COMMAND SETs/
EMULATE JDSU RX3
Default: JGR COMMAND SETs
In remote control, the MBR5 Meter may be
operated using the JGR command set, or
emulate a JDSU RX3 meter and accept the
RX3 commands.
INTERFACE: RS-‐232 / GPIB
Default: RS-‐232
The RS-‐232 selection prompts the Baud Rate
option.
The GPIB selection prompts the GPIB
address option
BAUD RATE: ____
Default: 9600
Choose a rate between 300 and 38400 kbit/s
GPIB ADDR: __
Default: 21
Select a GPIB address between 01 and 30
OPTION MENU
Compensate IL: ON/OFF
Default: ON
Monitors and compensates for internal
source drift
SETUPviaLOSS: ON/OFF
Default: ON
Adjusts the BR measurement by
compensating for loss at the connector.
The measurement is taken during IL
reference and offsets BR by twice this value.
MULTIDISPLAY: ON/OFF
Default: OFF
Enables Multidisplay mode to show 4
channels on the display at the same time
Save?
Quit?
Once changes are made the meter returns to
the measurement mode.
Compatibility with JDSU RM3 or RX3 Meters
The MBR5 can be set to emulate a JDSU RX3 backreflection meter via the
RS-‐232 or GPIB interface. The MBR5 will respond to RX3 commands if the user
sets the MBR5 into JDS Compatible Mode as follows:
1.
Access the User Menu as described in the User Menu Operation section.
2.
Select the “REMOTE MENU” option and use the softkey to change
“MBR5 Command Set” to “Emulate JDSU RX3”. Save this selection.