MBR5 User Manual
MBR5 Rev 007
58
[:POWER]:MODe?
Syntax
[:POWER]:MODe?
Function Returns the current operation mode.
Example
MOD?
returns: DUL.
[:POWER]:READ?
Syntax
[:POWER]:READ?
Function Returns the current reading (power or BR) of the meter.
Example
READ?
returns
-‐55.2 (the unit is in BR mode).
[:POWER]:READ:FULL?
Syntax
[:POWER]:READ:FULL?
Function Returns the current reading (power or BR) of the meter, the
current channel number, detector number and output
wavelength.
Example
READ:FULL?
returns
-‐55.2, 5, 0, 1310 (the unit is in BR mode).
[:POWER]:DETector[:NEXT]
Syntax
[:POWER]:DETector
<space>
[detector_number|MIN|MAX]
OR
[:POWER]:DETector[:NEXT]
Function Switches to the specified detector number
•
no parameter or [:NEXT]:
switches to the next detector
number
•
detector_number:
switches to the specified by the
parameter detector number
•
MIN or MAX:
switches to the first or last configured detector
number respectively.
Example
DET 2
switches to the detector number 2.