MBR5 User Manual
MBR5 Rev 007
56
*WAI Command
Syntax
*WAI
Function Prevents the meter from executing any further commands or
queries until all previously pending operations have been
completed.
There are no consequences to this command because all commands
are executed sequentially; therefore, any subsequent commands
are completed by the time this command is parsed.
Example
*WAI
SCPI Commands
[:SOURce]:WAVlength
Syntax
[:SOURce]:WAVlength
<space>
[wavelength|MIN|MAX|DEF]
OR
[:SOURce]:WAVlength[:NEXT]
Function Switches to a specified or next available wavelength.
•
no parameter or [:NEXT]:
switches to the next available
wavelength
•
wavelength:
switches to the specified wavelength in nm.
Other units can be specified
•
MIN, MAX or DEF:
switches to the first, last or default
wavelength respectively.
Example
WAV 1.55 um
switches the meter’s output to 1550 nm.
[:SOURce]:WAVlength?
Syntax
[:SOURce]:WAVlength?
<space>
[MIN|MAX|DEF]
Function Returns the current or specified output source wavelength in nm
•
No parameter:
returns the current wavelength
•
MIN, MAX or DEF:
returns the first, last or default
wavelength respectively.
Example
WAV? MIN
returns: 1310.