MBR5 User Manual
MBR5 Rev 007
40
Table 9: Front Panel Display Function
Display
Problem
Solution
Meter does not turn on Turn OFF the unit.
Connect the MBR5 meter to a
reliable power source, and wait a
few minutes.
Turn the unit ON.
INITIALIZING
Message is displayed for
a long time
Turn OFF the unit.
Connect the MBR5 meter to a
reliable power source, and wait a
few minutes.
Turn the unit ON.
*
Backreflection
measurement is getting
close to MBR5 meter
limit
Follow the setup and
measurement techniques
described in the “SETTING UP
THE METER FOR
BACKREFLECTION
MEASUREMENTS” section
<
BR range is very limited
Follow the setup and
measurement techniques
described in the “SETTING UP
THE METER FOR
BACKREFLECTION
MEASUREMENTS” section
TOO MUCH
LIGHT!
Dark current (ID) value
is too high
Attach the detector cap to the
detector, and press the ID key.
LASER POWER
LOW
Laser power is too low
Check the measurement setup
for possible source of light being
reflected back to the output port
of the source.
LASER POWER
HIGH
Laser power is too high
Check the measurement setup
for possible source of light being
reflected back to the output port
of the source.