To set a new target torque
(When % is the default limits mode)
1. Press and hold the limits key for 1 second.
(The current target torque shows with the first digit
blinking).
2. Use the increase and next key(s) to set the new target
torque (see “entering numbers” below).
3. Press the save key to accept the new target torque.
Note:
You will see the new target torque displayed in the
secondary set of digits when the limits mode is on. This
value will remain, even after turning the unit off, until you
either change the units of measure or plug in a new
transducer.
To recall the last cleared torque value
1. Press the recall key.
2. Press the clear key when finished viewing and return to
operation.
Note:
You can press the save key to save the recalled torque value
before pressing the clear key. This allows you to recall and
save a value that may have been accidentally cleared prior
to saving.
To save a torque value to memory
1. Press the save key to save the displayed torque value to
memory.
Note:
If you press the save key when zero is displayed you
will store the zero value in memory. Use the clear key if
you do not want to save to memory but want to clear the
display.
To view the attached transducer information
1. Press the transducer keys.
2. Press clear when finished viewing the data.
Note:
The range is in the upper right digits.
The MV/v (cal factor) is in the lower digits.
Smart Transducers will show “Auto Xducer”
Menu functions:
There are many settings that can be changed or
viewed with the menu function.
1. Activate the menu function by pressing the mode and
units key at the same time (the display will show set-up).
2. Press the NEXT key to step through the various
options. See bleow for the functions available
Press the save key to select the option that is displayed.
To recall memory:
Each time you press the save key to clear the
display you save the cleared value to memory. To
recall and view memory values do the following:
1. Press menu keys (set-up icon comes on).
2. Press the next key once or until the ”recall mem” icon is
on (if you miss keep pushing the next key until “recall
mem” shows.
3. Press save to select the recall memory option. The
number of data in memory now shows (up to approx. 1350
values can be saved).
4. Press the next key to scroll through and view the data
(press the increase key to view the previous torque data).
5. Press the clear key when finished (you will be returned
to normal operation mode).
Note: The data number is shown in the upper right digits.
IMPORANT!
If you fill up the memory you will hear a beep when you
press the save key during operation. Clear memory if this
occurs..