22
Troubleshooting:
Meter reads “OPEN” when turned on:
1. Transducer is not plugged in.
Plug in transducer
2. There is a significant load on the transducer
when the meter is powered on.
Remove load (such as hanging wrench).
3. The transducer or cable may have been damaged.
Return for repair.
Meter beeps when pressing save button:
1. There may be a load on the transducer when power
was turned on:
Remove load and press Zero Tare or
Remove load and turn meter off then on
1. Memory is full.
Clear Memory
Meter won’t show zero:
1. Recharbeable battery has reached re-charge limit:
Replace rechargeable 9 volt battery.
Use conventional Lion battery (IMPORTANT!
Do not use the power adapter when running
the meter with a conventional (non-rechargeable)
battery.
Specifications:
Standards: This tester when combined with JETCO torque transducers, is guaranteed to meet or exceed ANSI
B107.29 Electronic Tester, Hand Torque Tools standards.
Range and Display:
Direct reading triple numeric LCD display with large 4 digit
primary display (5/8” high) and two additional 4 digit displays.
Maximum reading 9999.
Floating Decimal Point.
Accuracy :
:
When used with:
Smart Bench top:+/- .5% IV from 10 – 100 % of transducer range +/- .1.
Smart Extension (IL): 1% IV from 10 – 100% of transducer range +/- .1.
Greater accuracy may be achieved with matched calibration.
Sampling Rate:
Approx. 8,000 samples per second.
Power:
Battery: 9 volt - ( rechargeable supplied for
Battery life:
Up to 20 hours with standard 9 volt battery under normal
use (time depends on use and battery quality).
Recharging Circuit:
Internal. 110 volt input, 9 volt 300 mA output, adapter supplied
standard (center positive 2.1 mm connector).
Meter Case:
Dimensions: 7.25” long x 3.25 “ wide x 1.25” thick (2.5” with stand)
Weight
Less than 8 ounces
Stand:
Folding and adjustable bench, wall or shelf stand supplied standard.
Memory:
Up to 1,350 torque data. Includes data number, torque value and units.
Communication:
Direct to serial computer or printer.
RS232C 4800 BPS, 8, 1, n.
Communication cable and adapter supplied standard.
Operating Modes:
Track, Peak and First Peak for calibration of micrometer (clicking) torque
wrenches.
Limits setting:
Tolerance values displayed on triple numeric display during operation.
Quick on/off with one key.
User selectable % (1-99%) with programmable target.
or
Programmable high and low values.
Tolerance prompting:
Triple lights and buzzer: Yellow (under), Green (within), Red (over)
Tolerance values displayed on triple numeric display during operation.
Display Clear Modes:
Automatic cycle complete.
Programmable timed auto reset from 1 – 9 seconds.
Manual Clear.
Filters:
5 user selectable filter settings: 170hz, 250 Hz, 500 Hz, 1500 Hz,
3600 Hz.
Transducer Selection:
Automatic “plug and play” with smart chip.
or manual “custom calibration”.
Operating Temperature: 40-110
°
F (72
°
F optimal operating temp.) If meter was stored in
cold environment allow internal temperature to stabilize prior
to use.
Storage Temperature:
0-140
°F
Degraded battery life:
23