EASYPOOL SMART 0 Operating Manual  |  17

Controller

4.4 Operation and menu structure

4.4.1 Operation

0
1
2

3
4
5

Fig. 4.15: View of TOPAX DX SMART housing with with operating panel.

Keys

Functions

Crossed arrow keys

       

•  Menu change-over in the menu "service";
•  Change-over between the individual numbers
•  Numerical values change, parameter adjust
Press the key to change from one menu item to 
the other. Numeric values are entered and modi-
fied continuously.

Key "OK" between 
the arrow keys

Input information are received and saved.
Successful saving is notified by a longer beep

Key 

a

: „ESC“

Exit the menu, one level up
Terminate input without storage

Key 

a

: „ESC“

(hold 5 seconds)

Return to the main menu

Keys 

b

 - 

f

The functions of the keys vary according to the 
menu and are displayed respectively.

TOPAX DX SMART key layout

If no key is operated, the controller returns automatically to the 
measuring mode after approx. 5 minutes. Changed parameters, 
which were not confirmed with the “OK” key, are not stored.
Exception: During the calibration and with the configuration no 
“time out” takes place.

NOTICE!

If no values are to be changed, "ESC" can be pressed 
and the selected menus left at any time. The control-
ler continues working with the old settings. Changed 
values  are  confirmed  with  the  “OK”  button.  The 
"clock" and "timer" functions are exceptions. Their 
modification does not require any confirmation.

General rules for entries:

The values can only be changed within the admissible value 
range.
The function of the key is signalled with a tone.
Some functions can be password-protected against unau-
thorised access (see chapter 4.1, menu 5.9 "Password").

NOTICE!

An auxiliary function is available for many menu items. 
In this case, the "Help" key is shown.

4.4.2 Menu structure

The  menus  are  as  a  result  shown  on  the  display  with  its  col-
oured presentation and high resolution in clear text and clearly 
arranged. As a result only the structure is described, not every 
detail. (for details of the service menu see chapter 4.1)

Standard display

If there is no entry made within 5 minutes, the device changes to 
the standard display (see Fig. 4.4..1).

7.04

0.39

26.3

mg/l

ºC

temperature

free Chlorine

pH value

eff. cl

0.30 mg/l

Topax DX SMART

14:28  31.03.2008

Y

Y

Fig. 4.4..1: Standard display

The  bars  below  the  digits  of  the  measurements  indicate  the 
strength  of  the  respective  output  signal.  The  display  colour 
changes, depending on the signal power, from green (0 ... 85%) 
to  orange  (85 ... 95%)  and  red  (95 ... 100%).  Blue  stands  for 
manual operation or basic load dosing.
By pressing a random key you go to the main menu (menu 1).

IMPORTANT!

All menus are given a number in the bottom left cor-
ner (see menu 1). See tab. 4.4.2.2.


Summary of Contents for EASYPOOL SMART 02

Page 1: ...02 Water sampling station for private swimming pool applications EN 01 Dosing Conveying Control Liquids Gases Systems Operating Manual Read this operating manual before start up To be retained for fu...

Page 2: ...nstallation and Commissioning 29 5 1 Tools required 29 5 2 Wall mount 29 5 3 Electrical connection 29 5 4 Connection to the pool water circuit 29 5 5 Fitting and calibrating sensors 31 5 6 Starting au...

Page 3: ...safety tips Basic safety precautions should always be followed when install ing and using this electrical equipment These precautions in clude the following WARNING 1 Read and follow all instructions...

Page 4: ...plicable regulations must be observed The operating manuals of prod ucts used in the system must be observed in par ticular the operating manual of the peristaltic pump which is included separately wi...

Page 5: ...re gauge e depending on model and pH electrode f The water sampling station is connected to the swimming pool water system All the items required for inclusion in the system and for disinfection are e...

Page 6: ...42401014 Yes 42401003 REDOX pH value Yes 42401004 Yes 42401000 Active oxygen 42401010 Flow through chlorine electrolysis 0 20 mA Chlorine pH value Yes Yes 42401001 Chlorine pH value Conductivity 0 20...

Page 7: ...rode Output signal 4 20 mA 0 20 or 0 60 mS cm Operating pressure 0 2 3 bar Operating temperature 5 40 C Water requirement approx 45 l h Weight approx 11 kg Hydraulic connections Clamped connection for...

Page 8: ...ity conductiv or inductive with separate measuring amplifier 20 mA type measuring range depending on type of measuring amplifier Digital inputs Early warning level input for disinfectant Warning level...

Page 9: ...hinge pivot from the equipment If both bolts are removed at the same time the upper part and the lower part of the box will fall loose NOTICE The equipment is to be opened only when plugged off the po...

Page 10: ...order no 78402 Terminal Function Analogue power outputs 0 4 20 mA see also chapter 4 10 1 Measurement value output 0 4 20 mA Disinfection 2 3 pH value 4 5 Temperature 6 7 Salt content 8 9 Programmed a...

Page 11: ...no 78404 optional Terminal Function 71 not used 72 73 74 20 mA passively without supply to the sensor Conductivity measurement 75 76 not used 77 78 outputs board order no 78399 Termi nal Output 81 Di...

Page 12: ...z Description Term Cable Remarks a Chlorine measuring cell CS120 51 A red RD platinum Pt 52 A Blue BU Copper Cu purple VT Silver Ag b pH single rod measuring cell 53 white WH Internal lead 54 black BK...

Page 13: ...ater contact 17 18 Description Term Cable Remarks f Disinfection peristaltic pump 88 Brown BN Relay 41 Blue BU N 39 Yellow green YE GN PE g Peristaltic pump pH 90 Brown BN Relay 41 Blue BU N 39 Yellow...

Page 14: ...sample water contact 17 18 Description Term Cable Remarks j Technoline SC 11 14 Flow through chlorine elec trolyses 1 power supply unit 2 control unit 3 cable A N 91190000 4 electrolytic cell 87 88 y...

Page 15: ...contact 17 18 Description Term Cable Remarks j Technoline SC 11 14 Flow through chlorine elec trolyses 1 power supply unit 2 control unit 3 cable A N 91190000 4 electrolytic cell 87 88 yellow YE Relay...

Page 16: ...ine dosing 90 Relay 42 N Description Terminal Remarks m Power supply Conductivity transmitter 45 L 41 N 39 PE m Conductivity measuring cell 78403 78399 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19...

Page 17: ...values are confirmed with the OK button The clock and timer functions are exceptions Their modification does not require any confirmation General rules for entries The values can only be changed with...

Page 18: ...values The input measuring values are examined for Free chlorine pH value REDOX potential Temperature Conductivity In doing so an important element is the calibration of the sen sors NOTICE When set u...

Page 19: ...ufficient reference value 2 only to check the chlorine content after optimization 4 5 2 Measuring input pH value The voltage signal transmitted by the combination electrode is proportional to the pH v...

Page 20: ...this solution a voltage of 12 mV can be displayed theoretically The physical value actually measured is however always different from the theroretical one When the physical value on the display become...

Page 21: ...calibrated the setting can be calculated from the following table The current measured mA is shown in the calibration window A hand held measuring device can be used for the reference measurement 4 6...

Page 22: ...ercentage value the actuator should reach Example a value of 80 operates the automatic controller to a maximum of 80 The algorithm of the automatic control is however computed further on 100 It is to...

Page 23: ...Tv caused by the differential rate drop to 0 even if the actual value keep deviating away from the default one The fact that the control causes the actual value to match the default one is caused main...

Page 24: ...chlorination For shock chlorination the free chlorine controller is switched to manual operation for an adjustable time and an adjustable dosing rate When the time has elapsed the controller automati...

Page 25: ...the logbook 5 7 Clock Setting of date and time 5 8 Timer Setting of control timer 5 9 Password Setting an access code for individual menus 5 10 Service code Setting a code for menu 5 service 5 11 DIN...

Page 26: ...s automatic changeover Menu 5 8 Timer Main menu menu 2 service timer In this menu you can adjust the timer for active oxygen dosing and allocate to individual week days Menu 5 9 and 5 10 code und Serv...

Page 27: ...be switched off in a specific way in menu 2 Service reset press the Switch off key All controllers are set to an output capacity of 0 and all unsaved data is saved to the memory card The controller c...

Page 28: ...MART the last one in the network The resistances are not active in its delivered state The 120 resistor is applied at the PC side via the interface converter The jumper slots are located in the front...

Page 29: ...th a spirit level After assembly put the water sampling station back in the wall bracket and attach the hinges 5 3 Electrical connection All work on the electrical installation must only be carried ou...

Page 30: ...pump and the injection nozzle is included in the supply The suction hose and pressure line should be routed without kinks Note the peristaltic pump entry and exit points Fig 5 4 2 2 Dosing line hose...

Page 31: ...des must be operated for approx 1 hour with pool water prior to calibration They must be removed for calibration To do this the ball valves at the input and output points of the water sampling station...

Page 32: ...oller description Provide the chemical supply for the pH adjustment and disinfec tion in turn by inserting the suction lines in turn into the vessel Note how the dosing starts each time via the perist...

Page 33: ...a container with water Store the electrodes in a frost free environment NOTICE Over long periods of inactivity an oxide layer forms on the copper electrode of the chlorine measuring cell This must be...

Page 34: ...ng of water guiding components Seal replacement Pump hose replacement Injection nozzle maintenance Sensor check Chlorine measuring cell maintenance The chlorine measuring cell is disassembled and visu...

Page 35: ...s and ball valves 38660 Copper electrode flat 22307 Platinum electrode spiral 22306 Silver electrode flat 35506 d Temperature sensor Pt 100 41100022 Table 8 3 Sensors spare parts for sensors and consu...

Page 36: ...drift too large Not activated No Change buffer solution or replace electrode Sensor or single rod measuring cell not compli ant to DIN Standards Check and eventually replace the input sensor Defective...

Page 37: ...he hand measurement is not affected Stabilise the pH reading through appropriate controller adjustment Potential carry over through electrical devices in the pool circuit Compensate potential insert m...

Page 38: ...ation 29 Installation height 29 Installing sensors 31 Insufficient sample water 30 Interruption in operation 33 L Language 27 M Main menu 18 Maintenance 34 Manual operation 24 Measured value alarms 24...

Page 39: ...EASYPOOL SMART 02 Operating Manual 39 Notes Notes...

Page 40: ...40 EASYPOOL SMART 02 Operating Manual Notes...

Page 41: ...dificaci n en la presente mercanc a sin nuestra previa autorizaci n esta declaraci n pierde su validez NL EU overeenstemmingsverklaring Ondergetekende Lutz Jesco GmbH Am Bostelberge 19 30900 Wedemark...

Page 42: ...erties We hereby certify that the product has been cleaned thoroughly inside and outside before returning that it is free of hazardous materi als i e chemical biological toxic flammable and radioactiv...

Page 43: ...scription of fault Type of fault 1 Mechanical fault Premature wear Wearing parts Breakage other damage Corrosion Damage in transit 2 Electrical fault Connections connectors or cables loose Operating c...

Page 44: ......

Reviews: