9 R
EMOTE
C
ONTROL
:DISP:OFF <NR1>
In order to save power, the display switches off automatically after a
predefined time of inactivity.
If an external power supply is connected, the display doesn’t switch off.
Valid values:
0 (15 sec)
1 (30 sec)
2 (1 min)
3 (2 min)
4 (10 min)
5 (Never)
Default:
5
:DISP:OFF?
Returns the time of inactivity after which the display automatically
switches off.
Response type:
<NR1>
See :DISP:OFF command for more information.
:ETH:MAC:ADDR?
Returns the actual MAC address of the instrument.
:IP:ADDR?
Returns the IP address which is currently valid for the instrument. In
static mode, this setup corresponds to IP:ADDR:STAT; in DHCP mode,
it is provided by the DHCP server.
This setup is read-only.
:IP:ADDR:STAT
<STRING>
Sets the static mode IP address.
:IP:ADDR:STAT?
Returns the static mode IP address.
Response type:
<STRING>
See :IP:ADDR:STAT command for more information.
:IP:GATE?
Returns the IP gateway address which is currently valid for the
instrument. In static mode, this setup corresponds to IP:GATE:STAT; in
DHCP mode, it is delivered by the DHCP server.
This setup is read-only.
:IP:GATE:STAT
<STRING>
Sets the static mode gateway address.
:IP:GATE:STAT?
Returns the static mode gateway address.
Response type:
<STRING>
See :IP:GATE:STAT command for more information.