UPDATE
Connector type
The connection is made with an universal push pull (UPP)
adapter for all 2.5 mm connectors (UPP adapter for 1.25 mm
ferrules are also available). The OLP-85 works with a .../PC and .../
APC connector as well.
Powermeter mode
Both modulated and unmodulated light signals can be
measured. The average power of modulated light signals is
displayed. The power meter is calibrated at 850 nm, 980 nm,
1310 nm, 1490 nm, 1550 nm and 1625 nm.
The “Auto-
λ
” function allows automatic wavelength detection.
This application requires a wavelength encoding light source,
e.g. one of the OLS-85 Optical Light Sources.