![JDS Uniphase OLP-85 User Manual Download Page 70](http://html1.mh-extra.com/html/jds-uniphase/olp-85/olp-85_user-manual_3631986070.webp)
70
OLP-85
9 R
EMOTE
C
ONTROL
C
OMMANDS
:SENS:MEAS:MODE?
Returns the active measurement mode.
Response type:
<NR1>
See :SENS:MEAS:MODE command for more information.
:SENS:POW:REF
<NR2>
Sets the reference level in dBm.
Default:
0
:SENS:POW:REF?
Returns the reference level in dBm.
Response type:
<NR2>
See :SENS:POW:REF command for more information.
:SENS:POW:UNIT
<NR1>
Select the unit for the absolute level.
Valid values:
0 (dBm)
1 (Watt)
Default:
0
:SENS:POW:UNIT?
Returns the unit for the absolute level.
Response type:
<NR1>
See :SENS:POW:UNIT command for more information.
:SENS:THRS:PASS
Sets the threshold value “PASS” for the actual wavelength/unit dBm.
Default:
0
1)
:SENS:THRS:PASS?
Returns the threshold value “PASS” in dBm.
Response type:
<NR2>
See :SENS:THRS:PASS command for more information.
:SENS:WLEN?
The actual “active” wavelength. This setup is driven by other setups and
results. (read only)
:SENS:WLEN:TBL
Table of selectable wavelength.
:SENS:WLEN:TBL?
Returns the table of selectable wavelength.
See :SENS:WLEN:TBL command for more information.
:SENS:WLEN:TBL
:LMT
Is a child map of :SENS:WLEN:TBL containing the threshold limit values
(unit dBm) for all selectable wavelength. Adding or deleting entries in
:SENS:WLEN:TBL will automatically update this map.
1)
Command string
Parameter type / Response type / Unit / Info
Summary of Contents for OLP-85