![JDS Uniphase OLP-85 User Manual Download Page 69](http://html1.mh-extra.com/html/jds-uniphase/olp-85/olp-85_user-manual_3631986069.webp)
OLP-85
69
C
OMMANDS
9 R
EMOTE
C
ONTROL
Setups for the Powermeter application
Command string
Parameter type / Response type / Unit / Info
:APP:ID:VSBL
<STRING>
Makes the specified application visible (i.e. brings the specified
application to foreground).
Notes:
Full functionality is only available for the active
application, so it may be important to know which
application is active.
Valid values:
All application IDs returned by :APPM:LIST?.
:APP:ID:VSBL?
Returns the name of measurement application which is currently visible
on the display.
Response type:
<STRING>
See :APP:ID:VSBL command for more information.
:AUTO:LMBD
<BOOLEAN>
Determines whether the instrument recognizes Auto
λ
modulation
frequencies.
Default:
TRUE
:AUTO:LMBD?
Returns TRUE if the instrument recognize Auto
λ
modulation
frequencies.
Response type:
<BOOLEAN>
See :AUTO:LMBD for more information.
:AUTO:LMBD:DTCT?
This (read-only) setup is set to true when the instrument detects an
automatic
λ
signal and auto lambda functionality is enabled.
:EVAL:ENAB
Enables or disables the PASS/FAIL evaluation.
:EVAL:ENAB?
Returns the current state of :EVAL:ENAB.
:MEAS:ACT?
<BOOLEAN>
Shows the status of the measurement. Valid results are available if this
setup is TRUE.
:SENS:MEAS:MODE
<NR1>
Selects the measurement mode.
Valid values:
0 (Absolute)
1 (Relative)
2 (PassFail)
Default:
0