![JDS Uniphase OLP-85 User Manual Download Page 73](http://html1.mh-extra.com/html/jds-uniphase/olp-85/olp-85_user-manual_3631986073.webp)
OLP-85
73
C
OMMANDS
9 R
EMOTE
C
ONTROL
Extended commands for the Powermeter application
Command string
Parameter type / Response type / Unit / Info
:SENS:ABS:TO:REF
Executing this command saves the current optical power as reference
value.
See :SENS:POW:REF for further information.
:STOR:ALL:DEL
Deletes all measurement entries and groups. Creates a new empty
“default” group.
:STOR:GRP:NAME is set to default.
:STOR:DEL
<STRING>,
<STRING>,
<STRING>
Deletes a dedicated measurement result specified by group (string1),
fiber identifier (string2) and internal identifier (string3).
:STOR:DEL:MLTP
<STRING>,
<STRING>,
<STRING>
Deletes multiple measurement results specified by group (string1), a
list of comma separated fiber identifiers (string2) and a list of comma
separated internal identifiers (string3).
e.g. :STOR:DEL:MLTP “TEST”,”ID1,ID1,ID2”,”00005,00009,00001”.
:STOR:GRP:CREA
<STRING>
Creates an empty storage group.
:STOR:GRP:DEL
<STRING>
Deletes the storage group specified by string.
If setup :STOR:GRP:NAME contains this group, it will be changed to
default.
If the “default” group is deleted, only the content of the group is
deleted.
:STOR:RD?
<STRING>,
<STRING>,
<STRING>
Reads out a dedicated measurement result specified by group (string1),
fiber identifier (string2) and internal identifier (string3).
:STOR:WR
<STRING>,
<STRING>
Writes a dedicated measurement result specified by group (string1)
and fiber identifier (string2). If the given group doesn’t exist, it will be
created.
This command creates a dedicated file for each existing wavelength. To
create unique file names, this function automatically adds an internal
sequence number.
Summary of Contents for OLP-85