REF. 53794
8
BEFORE TAKING A MEASUREMENT
1. Try the device by yourself first to know how to use it.
2. It must be placed in a general environment of room
temperature of 16ºC-35ºC for at least 30 minutes
before use to avoid incorrect measurement results.
3. The person measured should be kept at room
temperature for at least 20 minutes to make body
temperature balanced.
4. After strenuous exercise, you should rest for at least
30 minutes before taking measurement.
5. Wipe the skin dry and push aside hair strands
beforecommencing.
6. When repeated measurements are made
continuously, move away the probe between eachmea-
surement and rest for at least 5s before making the
next to obtain the most correct result.
7. The normal reading value is about 35.5ºC 37.8ºC. 8.
The user should take more temperature to know his or
her own health temperature. Pay attention to physical
changas or consult your doctor if there is any abnorma-
lity.
8. Check the probe is clean before each measurement
to ensure accuracy.
RECALLING READINGS FROM MEMORY
1. Ensure the thermometer is switched off.
2. Press the “ON/OFF/MEASURE” button for 4
seconds to put the device on memory mode. The last
reading will be displayed.
3. Press and release the “ON/OFF/MEASURE” button
again to display the next reading in a sequence from
the last reading to the first reading.
4. The thermometer has memory to recall the last 32
readings.
5. The thermometer will automatically turn off if it idles
for more than 12 seconds.
CHANGING TEMPERATURE SCALES
1. Ensure the thermometer is switched off.
2. Press the “ON/OFF/MEASURE” button for 8
seconds to bring it into temperature mode. The current
temperature unit will be displayed.
3. Release and press the ON/OFF/MEASURE button
again to toggle and select between ºC and ºF.
4. The thermometer will switch off automatically if it
idles for more than 4 seconds and will display the last
reading
WARNING!
• Do not use the thermometer under extreme tempera-
tures (below 16ºC/60.8ºF or over 35ºC/95ºF) or
humidity extremes (below 15%RH or over 80%RH).
• Do not use the thermometer under extreme tempera-
tures (below -20ºC/-4ºF or over 55ºC/131ºF) or
humidity extremes (below 15%RH or over 93%RH).
• If temperature probe has been shocked and lux,
contact your retailer immediately.
• lt is dangerous for patients to perform a self-evalua-
tion and self-treatment based on the measuring results.
Be sure to follow doctor’s instruction.
• Do not touch or blow infrared sensor.
• Clean a polluted infrared sensor with a soft dry cloth in
a gentle manner. Cleaning with toilet tissue or paper
towel may scratch the infrared sensor, causing
inaccuracy.
• Install the battery in the right position according to the
manual.
• Do not immerse the thermometer in liquids, this
device is not waterproof.
• Do not use a mobile phone nearby when the thermo-
meter is working.
• If the probe glass is damaged, stop using it and
contact your retailer immediately.
• Do not modify the device.
• The device is not repairable and contains no user
serviceable parts.
• The device requires no calibration.
• If the device is modified, appropriate inspection and
testing must be conducted to ensure continued safe
use of the equipment.
• The operator shall not touch the thermometer and the
patient simultaneously.
• Keep the device out of children’s reach.
• For temperature difference between storage area and
measurement site, condition the thermometer for about
30 minutes in room temperature.
• Ensure that children do not use the instrument
unsupervised. Some parts are small enought to be
swallowed.
• No servicing/maintance white the MW equipment in
use.
• The patient can measure data and change battery
under normal circumstances,
• The patient can maintain the device according to the
user manual.
ERROR MESSAGES
ELECTROMAGNELIC COMPATIBILAY
1. This product needs special precautions regarding
electromagnetic compatibility (EMC) and needs to be
installed and put into service according to the EMC
inforrnation provided, and this unifican be affected by
portable and mobile radio frequency (RF) communica-
tions equipment
2. This machine must not be used adjacent to or
stacked with other equipment, and if an adjacent or
stacked unit is necessary, this device should be
checked to verify normal operation in the configuration
to be used.
3. Caution:This unit has been thoroughly tested and
inspectedfor its proper performance and operation.
4. Carbon: This machine must not be used adjacent to
or stacked with other equipment, and if an adjacent or
stacked unit is necessary, this device should be
checked to verify normal operation in the configuration
to be used.
EXPLANATION OF MARKS OR SYMBOLS
The following symbols may appear on the manual,
Infrared Thermometer and its accessories.
MAINTENANCE, STORAGE AND CALIBRATION
Maintenance
1. Remove any stains on the body with a soft and dry
cloth.
2. Clean the probe tip as follows: very gently wipe the
surface with a cotton swab or soft cloth moistened with
clinical alcohol, after the alcohol has completely dried
out, you can use the thermometer. If the probe tip is
damaged, please contact the maintenance center.
3. Don’t wash with water or detergent containing
abrasive or benzene.
4. Do not immerse in liquids.
Storage
1. The device must not be stored or used at an excessi-
vely high or low temperature or humidity, in sunlight, in
association with un electrical current or in dusty
locations.
2. Remove the battery if non-use for a long time.
Calibration
This thermometer is calibrated at the time of manufac-
tured. If the thermometer is used according to the
instruction, periodic recalibration is not required. If at
any time you question the accuracy of the measure-
ment, contact the retailer immediately. Don’t attempt to
modify or reassemble the thermometer.
WARRANTY
1 year warranty from the date of purchase, excluding
errors made by the user below:
1. Error produced by unauthorized disassembly and
modification.
2. Error produced by a fall during use or transport.
3. Error produced by use far from the instructions
described in the manual.
INCLUDED IN DELIVERY
• Infrared thermometer
• Manual
STANDARD
This product is made under the ISO80601-2-56.
Error
message
Problem
Solution
Hi
Temperature is
higher than 42.2°C
(108°F).
Operate the thermometer
temperature ranges. In the
event of a repeated error
message, contact your
retailer.
Lo
Temperature is
lower than 32 °C
(89.6 °F).
Operate the thermometer
temperature ranges. In the
event of a repeated error
message, contact your
retailer.
Err
Sensor not welded. Contact your retailer.
Low power.
Replace new batteries.
ErH
The temperature
of the measuring
environment is too
high.
Decrease the temperature
of the environment, keep it
within 16°C-35°C.
ErL
The temperature
of the measuring
environment is
too low.
Increase the temperature
of the environment, keep it
within 16°C-35°C.