![Jasco V-530 Instruction Manual Download Page 80](http://html1.mh-extra.com/html/jasco/v-530/v-530_instruction-manual_3423385080.webp)
67
7.1.3 [P aram eter...]
S ets param eters, and saves them to the hard disk.
C lick [M easurem ent] - [P aram eter...] to display the dialog box show n below .
F igure 7.3 [param eters] dialog box
[P hotom etric M ode]
S ets the photom etric m ode.
S electable range:
A bs:
A bsorbance m easurem ent
% T :
T ransm ittance m easurem ent
% R :
R eflectance m easurem ent
S am ple:
S ingle-beam m easurem ent on the sam ple beam side
R eference:
S ingle-beam m easurem ent on the reference beam side
N ote: For m odels V -550/560/570, w hen [S am ple] or [R eference] is selected, [P M T voltage] is added to the left
side. Input the voltage to be applied to the photom ultiplier. The input range is 0 to 1000 V .
[R esponse]
R esponse by sim ple m oving average. S electable range varies according
to the m odel.
V -530:
Q uick:
M oving average during about 0.03 sec
Fast:
M oving average during about 0.25 sec
M edium :
M oving average during about 1 sec
V -550/560/570:
Q uick:
M oving average during about 0.03 sec
Fast:
M oving average during about 0.25 sec
M edium :
M oving average during about 1 sec
S low :
M oving average during about 4 sec
[B and w idth]
S pectral bandw idth. S electable range varies according to the m odel.
V -530:
F ixed at 2 nm
V -550/560:
0.1, 0.2, 0.5, 1, 2, 5, 10 nm
V -570:
0.1, 0.2, 0.5, 1, 2, 5, 10 nm
0.4, 0.8, 2, 4, 8, 20, 40 nm (near infrared
region)