[M ethod]
N o. of w avelengths used in quantitative analysis. S elects optim um
num ber. of w avelengths (1, 2, or 3) according to the sam ple condition.
1-w avelength:
F or com m on solution sam ple. S ee F ig. 4.13 (1).
2-w avelength:
F or baseline correction. S ee F ig. 4.13 (2).
3-w avelength:
F or baseline correction. S ee F ig. 4.13 (3).
In [3-w avelength], the absorbance value is obtained from the follow ing
equation:
D
Abs
E
WL
WL
E
WL
WL E
WL
WL
=
-
-
×
+
-
×
-
( )
|
| ( ) |
| ( )
|
|
1
1
2
3
3
1
2
3
2
WL 1
WL 1
WL 2
D Abs
D Abs
WL 1
WL 2
WL 3
E(2)
E(1)
E(3)
(1)
(2)
(3)
1-wavelength
2-wavelength
3-wavelength
(Peak)
(Base 1) (Peak)
(Peak)
(Base 1)
(Base2)
F igure 4.12 Q uantitative analysis m ethod according to num ber of w avelengths.
[P eak]
P eak w avelength
[B ase 1]
B ase 1 w avelength
[B ase 2]
B ase 2 w avelength
T he input range varies according to the m odel.
V -530:
190.0 to 1100.0 nm
V -550/560:
190.0 to 900.0 nm
V -570:
190.0 to 2500.0 nm
[S am ple N o.]
S ets the sam ple num ber for m easurem ent. S am ple num ber increases
increm entally by one w ith each subsequent m easurem ent.
[N o. of C ycle]
S ets how m any tim es each sam ple is m easured. If 2 or m ore
m easurem ents are set, the [C ycle T im e] field appears.
[C ycle T im e]
S ets the tim e in seconds betw een m easurem ents. If the cycle tim e is
shorter than the m easurem ent tim e, the next m easurem ent starts
im m ediately.
Input range: 0 to 15000 sec.
<O K >
T ransfers the m easurem ent param eters to the spectrophotom eter. T he
[C alibrate C urve P aram eters] dialog box appears.