Page 3
Jandy
®
UV Medium Pressure | Installation & Operation Manual
ENGLISH
Section 1. Important Safety Instructions
READ AND FOLLOW ALL INSTRUCTIONS
1.1 Safety Instructions
WARNING
When installing and using electrical equipment, basic safety precautions should always be followed, including the following:
A yellow-green wire connector marked
*
is provided on this unit to connect a minimum No. 8 AWG for US (UL) and a No. 6 AWG for Canada (CSA)
solid copper conductor between this unit and any metal equipment, metal enclosures of electrical equipment, metal water pipe or conduit within
5 feet (1.5 m) of the unit.
Two bonding lugs marked
*
are provided on the end flanges of the UV chamber suitable for No. 8 AWG (US) and No. 6 AWG (Canada) and secured
to the chamber by paint breaking washer and nut.
* IEC 60417, symbol 5019
WARNING
This product must be connected to a circuit protected by a ground fault circuit interrupter.
Pool end use. Do not use this system to a different application for which it was designed.
The equipment should be installed and handled by truly qualified people.
Current electrical and accident prevention regulations should be followed.
Under no circumstances will the manufacturer be held responsible for the assembly, installation or start-up, nor any handling or fitting of components
unless they are carried out on its premises.
WARNING
Check all the electrical connectors are well tightened to avoid false contacts and their consequent overheating.
Install the control panel so that the cooling grids are not obstructed.
For indoor use only. This unit is not intended for outdoor use.
Prior to the installation or replacement of any system component make sure it has been previously disconnected from the mains, and there is no water
flow through it. Use only spare parts supplied by Jandy.
WARNING
Never remove the lock nut of the quartz sleeve when the water is recirculating through the UV reactor as it could be expelled and causing damage.
The UV light generated by this equipment can cause serious damage if the eyes or skin are exposed directly to the lamp. Never connect the system
when the lamp is out of the reactor.
Do not handle the UV lamp until completely cold.
Always handle the UV lamp with gloves, as fat and other impurities deposited on the surface may reduce its performance and durability. In case you
have to clean the lamp surface use a soft cloth soaked with alcohol.