Integral
Operation Function
Copyright © Itech Electronics Co., Ltd.
26
Integration time
Displayed as Time, in the format of hhhh:mm:ss.
Specification
Save in case of power failure
The integral result can be kept in the memory even in case of power failure
during operation of the integral function. When the power supply is recovered
and the integral function is disabled, the integral result before power failure is
displayed. When the power supply is recovered and the integral is rest, the
integral function is enabled again.
Display resolution
The maximum display resolution of the integral value is 99999. When the
integral value reaches 100000, the decimal point will move automatically. For
example, when 0.01mwh is added to 999.99mwh, 1.0000wh will be displayed.
Display in case of overflow
When the integral value meets the following overflow conditions, the integral
function will be disabled, and the integral time and integral value at this point will
be kept.
The integral time reaches the maximum value (10000 hours).
The integral values of WP, q and WPAV reach the maximum/minimum
display integral values.
Maximum/minimum display integral value
Active power integral (WP): ±99999 Mwh
Current integral (q): ±99999 MAh
Average power integral (WPAV): ±99999 Mw
Restricted implementation
Settings of the following functions cannot be changed during integral operation.
Function
Integration status
Integral
resetting
Current
integral status
Integral
interruption
Functions related to settings of measurement parameters
Wiring
Executable
Unenforceable
Unenforceable
Measurement range
Executable
Unenforceable
Unenforceable
Filter
Executable
Unenforceable
Unenforceable
Averaging function
Executable
Unenforceable
Unenforceable
Synchronization
Executable
Unenforceable
Unenforceable
Data updating rate
Executable
Unenforceable
Unenforceable
Integral function parameter setting
Integral mode
Executable
Unenforceable
Unenforceable
Integral timer
Executable
Unenforceable
Unenforceable
Integral operation
Integral start
Executable
Unenforceable
Executable
Integral stop
Unenforceable
Executable
Unenforceable
Integral resetting
Executable
Unenforceable
Executable
Other operations
Save operation
Executable
Executable
Executable
Hold operation
Executable
Executable
Executable
Single measurement
operation
Executable
Executable
Executable