Waveform
Display Function
Copyright © Itech Electronics Co., Ltd.
19
Parameter name
Parameter descriptions
KNOB SEL
(U/I/TL/TD/Td)
Knob selection: the following variables can be
adjusted by rotating the knob: vertical voltage
position/vertical current position/trigger
level/trigger delay/level calibration.
AUTO
Automatic adjustment key: when the soft key
corresponding to this parameter is pressed, the
power meter will automatically calibrate the input
signal to display the best effect of the input signal.
TRIG SET
Trigger setting
MEASURE SET
Measurement setting
Introduction of waveform display interface
The trigger status is described as follows:
Trigger
status
Instruction
Auto
When the trigger mode is set as Auto, the trigger status Auto
will be displayed after triggering.
Auto?
When the trigger mode is set as Auto, the trigger status will be
Auto in the case of no triggering?
Trig
When the trigger mode is set as Normal, the trigger status Trig
will be displayed after triggering.
Trig?
When the trigger mode is set as Normal, the trigger status will
be Trig in the case of no triggering?
Stop
When the “Stop” soft key in the waveform display interface is
pressed, the trigger status Stop will be displayed.
Vertical calibration
When the crest factor CF is 3, the selected voltage range and current range will