![ITech IT8600 Series Programming Manual Download Page 67](http://html1.mh-extra.com/html/itech/it8600-series/it8600-series_programming-manual_2095285067.webp)
Calibration Commands
Copyright © Itech Electronic Co., Ltd.
56
Parameter:
P1|P2|P3|P4
CALibrate:VOLTage:METer
This command is used to calibrate the measurement voltage level.
Command syntax:
CALibrate:VOLTage:METer[:LEVel] <NRf>
Parameter:
NRf, external reading
Unit:
V
CALibrate:CURRent:METer:POINt
This command is used to calibrate the measurement current point.
Command syntax:
CALibrate:CURRent:METer:POINt <point>
Parameter:
P1|P2|P3|P4
CALIbrate:CURRent:METer
This command is used to calibrate the measurement current level.
Command syntax:
CALibrate:CURRent:METer[:LEVel] <NRf>
Parameter:
NRf
CALibrate:RESistance:POINt
This command is used to calibrate the measurement resistance point.
Command syntax:
CALibrate:RESistance:POINt <point>
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com