Calibration Commands
Copyright © Itech Electronic Co., Ltd.
54
Parameter:
None
For example: CAL:INIT
Relevant Command:
CAL:STAT CAL:INIT
CALibrate:SAVe
This command is used to save calibration coefficients in nonvolatile memory.
Command syntax:
CALibrate[:USER]:SAVe
Parameter:
None
CALibrate:VOLTage:POINt
This command can only be used in calibration mode. It is used to set the
calibration points of constant voltage mode. P1, P2 is used in low voltage meter
range, P3, P4 is used in high voltage meter range. It can use calibrate voltage
source and voltage meter.
Command syntax:
CALibrate:VOLTage:POINt <point>
Parameter:
P1 | P2 | P3 | P4
Example:CAL:VOLT:POIN P2
Relevant Command:
CAL:STAT CAL:SAV
CALibrate:VOLTage
This command is only used in calibration mode. It enters a calibration voltage
value that you obtain by reading an external meter. You must first select a
calibration level (with CALibrate:VOLTage:POINt) for the value being entered.
These constants are not stored in nonvolatile memory until they are saved with
CALibrate:SAVE.
Command syntax:
CALibrate:VOLTage[:LEVel] <NRf>