![ITech IT8600 Series Programming Manual Download Page 36](http://html1.mh-extra.com/html/itech/it8600-series/it8600-series_programming-manual_2095285036.webp)
Load Commands
Copyright © Itech Electronic Co., Ltd.
25
protection.
Command syntax:
[SOURce:]CURRent:PROTection:DELay <NRf+>
Parameter:
MINimum|MAXimum|DEFault or NRf
Unit:
s
Reset value:
MAXimum
Query Syntax:
[SOURce:]CURRent:PROTection:DELay? [MINimum|MAXimum]
Returned value:
NR3
CURRent:PROTection:STATe
This command is used to enable or disable the overcurrent protection of
electronic load.
Command syntax:
[SOURce:]CURRent:PROTection:STATe <bool>
Parameter:
0|1|OFF|ON
Query Syntax:
[SOURce:]CURRent:PROTection:STATe?
Returned value:
0|1
POWer
This command is used to read power value of Watts as a unit.
Command syntax:
[SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude] <NRf+>
ООО "Техэнком" Контрольно-измерительные приборы и оборудование www.tehencom.com