9.7. ModuleControl (read access)
:READ:MOD
ule
:CONTR
ol
? UI2
Bit15
Bit14
Bit13
Bit12
Bit11
Bit10
Bit09
Bit08
Reserved
Set Kill Enable Reserved
Set Fine
Adjustment
Set Big Endian Reserved
Reserved
Reserved
Bit07
Bit06
Bit05
Bit04
Bit03
Bit02
Bit01
Bit00
Reserved
Do Clear
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Status Bit
Description
Set Kill Enable
Kill function
Set Fine Adjustment
Switch ON of fine adjustment
Set Big Endian
Order of bytes in word: 0 = Little Endian (INTEL); 1 = Big Endian (MOTOROLA)
Do Clear
Hardware Clear Kill signal and clear all event signals of the module and the channels
Reserved
Control Bit
Description, when the Bit is “0”
Description, when the Bit is “1”
Set Kill Enable
kill function disable
kill function enable
Set Fine Adjustment
fine adjustment OFF
fine adjustment ON
Set Big Endian
big endian (MOTOROLA format)
Do Clear
no action
Hardware Clear Kill signal and clear all event
signals of the module and the channels
EHQ High Voltage Power Supply series | Last changed on: 11.05.2020 |
www.iseg-hv.com
38/46