9.6. Module EventStatus (read access)
:READ:MOD
ule
:EV
ent
:STAT
us
?
UI2
Bit15
Bit14
Bit13
Bit12
Bit11
Bit10
Bit09
Bit08
Reserved
Event
Temperature
Not Good
Event Supply
Not Good
Reserved
Reserved
Event Safety
Loop Not
Good
Reserved
Reserved
Bit07
Bit06
Bit05
Bit04
Bit03
Bit02
Bit01
Bit00
Reserved
Reserved
Reserved
Reserved
Event Service
Reserved
Reserved
Reserved
Status Bit
Description
Event Temperature Not Good
Temperature is above 55°C
Event Supply Not Good
at least one of the supplies is not good
Event Safety Loop Not Good
Safety loop is open
Event Service
A hardware failure of the HV module has been detected. The HV is switched off without the
possibility to switch on again. Please
the iseg Spzialelektronik GmbH.