Cleaning Instructions
CAUTION
Due to the close proximity of electrical power and equipment, flammable cleaners should never be used to clean
these products!
The surface materials of the unit are primarily powder-coated aluminum and are durable and easy to maintain,
however they can stain and discolor, so test any cleaners in an inconspicuous place before using.
Do not allow any liquids to enter the unit, drip down the monitor or accumulate on any surface.
Please refer to the respective Materials Safety Data Sheets (MSDS) for detailed descriptions for each product
from its manufacturer.
Never use steel wool, Scotch-Brite™ or other abrasive materials to clean the product.
Use extreme caution when cleaning the camera, as it is delicate and easily broken.
Use extreme caution when cleaning a display monitor, as they are easily damaged if too much pressure is
applied.
General Procedure
1.
Verify the system is unplugged from the AC Power outlet before cleaning.
2.
Use a soft, clean microfiber cloth or manufacturer supplied disposable cloth for all applications, particularly
when cleaning lenses and monitors. Do not spray liquids directly on the surface.
3.
Utilize appropriate cleaners for the surface being cleaned.
4.
Allow equipment to fully dry prior to plugging into a power source.
5.
To facilitate an effective infection control program and ensure proper performance, routinely clean, disinfect,
and maintain products in accordance with approved procedures. Specifically, the hospital’s Infection Control
Administrator should be consulted for cleaning procedures and processes.
Suggested chemical cleaners/disinfectants/solutions for CLINiC and MedView:
Chassis cleaning
o
Non Abrasive Soap/Detergent: Generally, water and mild non-abrasive soap/detergent or isopropyl
alcohol can be used routinely on CLINiC or MedView products to maintain proper cleanliness.
o
Where infection control is required
A 10% or less bleach solution can be used to disinfect. Remove residue using a clean damp (water)
cloth.
Branded chemical disinfectant products (test specific product on a sample surface before general
use)
Metrex CaviWipes
Clorox Germicidal Wipes
Display monitor LCD panel and camera body (not the lens)
o
Do not use any of the following chemicals or any solutions that contain: chlorine (bleach), acetone,
peroxides, ammonia, ethyl alcohol, benzene, toluene, ethyl acid, or methyl chloride.
o
Branded, ammonia-free LCD cleaning products
Zeiss Pre-Moistened Lens Cleaning Wipes
CloroxPro Clean Screen Wipes
o
Up to 50:50 isopropyl alcohol to distilled water mixture for general cleaning, using soft microfiber cloth
o
Use 70:30 isopropyl alcohol and distilled water mixture for infection control, using soft microfiber cloth
Camera Lens
o
Use only branded, ammonia-free cleaning wipes specifically designed for lens cleaning
Zeiss Pre-Moistened Lens Cleaning Wipes