| LaserNet software
84
YLS-K
Channel number
(Option depends on configuration)
These fields display the current coding for each optical channel of the
beam switch. The coding corresponds to the optical path that must be
requested by the system owner’s PLC unit. If the first optical channel of
the beam switch was coded with the value 10, the PLC unit must set op-
tical path number 10. The coding is performed in the laser.
Laser number
The current coding of the laser is shown here. The encoding corre-
sponds to the laser number which must be requested from the opera-
tor-side PLC unit if the check box
Check
in the
Laser number
group
box has been activated. If the laser is coded with value 6, the PLC unit
was set laser number 6. The coding is performed in the laser itself.
Check
If this option is activated, the laser control checks whether the right
laser number is sent by the system owner’s PLC unit. If the laser num-
ber is not correct, the laser reports a failure.
Laser program number
If you activate the option
Laser program number
, the input signal of the
external control Laser program number is immediately reported back
to the external control. Otherwise, the laser program number is only re-
ported after receipt of the signal
Program Start
along with the execu-
tion of the program.
Update configuration file
Clicking this button opens a file browsing window with which you can
upload an update file (dbm format) for configuration of your Profinet
slaves.
This update file can be obtained from IPG Laser GmbH upon request.
Advanced settings
Clicking this button opens an input window in which you can change
various parameters depending on the available fieldbus interface.