153
YLS-K
Possible
causes
The cause is an internal error in the combiner module.
Analysis / trou-
bleshooting
ð
Using the guide laser, check whether the internal optics of the
laser is damaged. To do this, switch on the guide laser and check
whether this is visible after the optics head.
The guide laser should show a round shape in the area of the focus. If the guide
laser is not visible or is poorly visible, this is an indication of damage to the internal
optic.
ð
Follow the steps described in section
for resetting critical errors.
Message
Initialization error
Type
Fault
Reset
Manual
Description
The alarm
Initialization failure
indicates that the operating status
Laser
ready
is not achieved. The status
Laser failure
is activated.
Possible
causes
•
Firmware error
Fault diagnos-
tics / trou-
bleshooting
ð
Contact the responsible IPG Service department.
Message
Laser temperature too low
Type
Fault
Resetting
Automatic
Description
The alarm
Laser temperature too low
is activated when the average
module temperature of all modules is below 18 °C. The operating status
Laser ready
is deactivated, and the status
Laser error
is activated.
Possible
causes
•
The parameterization of the laser chiller is faulty.
•
The tank heater of the laser chiller is defective or not switched
on.
•
The laser stood for a prolonged period of time in a cold environ-
ment while switched off.
Analysis / trou-
bleshooting
ð
Check the parameterization of the laser chiller.
ð
Check the tank heater of the laser chiller.
ð
Switch on the laser chiller. The heating elements of the laser
chiller heat the water to the necessary temperatures.
IPG chiller:
ð
Using the
Chiller
tab, determine what type of fault this is.
ð
See
Messages of the Chiller tab
or read the operating manual of
the laser chiller for a detailed fault analysis.
| Appendix