115
YLS-K
1
2
Figure 69:
Unplugging fiber connectors (here: a HLC-8 fiber plug into an FCH-8 adapter)
9.
Immediately seal the optics with the cleaned protective cap.
10.
Place the fiber connector in the holder on the microscope (see fig-
ure Microscope).
11.
Switch the microscope's light source on and point this towards
the surface to be cleaned.
Figure 70:
Microscope with fiber connector
Item Designation
Item Designation
1
Movable light source
4
Quartz block
2
Microscope
5
Mount for the fiber connector
3
Pan head
6
Fiber connector
| Maintenance