4-18 Edit Menu
889691
eZ-Analyst
TEDS Info
When IEPE is selected you have the option to use TEDS. You can then hold down
the <Ctrl> key and click the mouse button to bring up TEDS information as
indicated in the following screen image.
Example of TEDS Information
Note
: the Preferences Tab contains a checkbox with the associated text:
“Check for TEDS sensors on Dynamic Input Channels.”
If checked, every time eZ-Analyst is started, every channel [on each connected
WBK18 or ZonicBook main channel] is scanned for the presence of a TEDS
compliant accelerometer. This scanning adds time to the eZ-Analyst startup.
See Tip
.
TIP
: If your acquisition system doesn't have [or is not making use of] TEDS
accelerometers, then leave the associated checkbox in the Preferences Tab
“unchecked.” This will allow for a faster startup of eZ-Analyst.
Coupling
AC or DC coupling can be set on an individual channel basis from the Input
Channels tab. Coupling is available for analog channels (CH1 through CH8) and
for compatible expansion channels [e.g., from a WBK18]. AC Coupling can be set
to a high pass filter value of 0.1 Hz or 1 Hz. The value is selected in the Edit
Menu’s Analyzer Tab (see page 4-4). Note that when DC Coupling is selected the
high pass filter is bypassed.
When the Coupling function is not available to a channel, or not used by a
channel, a dash appears in that channel’s Coupling column.