3-8 Menus
989090
eZ-Analyst
Calibrator
EU
This field is used to set the Calibrator’s Engineering Units.
Measured
mV
Displays the measured value of the channel in milli-volts.
Transducer
mV/EU
The Transducer mV/EU value can be entered manually, or by the
measurement process. Highlight a cell, or cells, in the grid before typing a
value in the data entry box. Then press the <OK> button to accept the
value. All other fields linked to this value are updated when the value is
accepted.
mV/EU= Volts/Units * Transducer Gain
EU Offset
EU Offset is used for DC signal compensation. Offset is added to, or
subtracted from the measured EU value.
Transducer
Gain
Transducer Gain is used as auxiliary scaling to compensate for the
transducer amplifier gain. Gain is a multiplied function.
Calibration
Date
This column displays the date and time of the last calibration. If any
channel value changes, even if the original number is restored, the
calibration date and time are automatically removed for that channel.
Start
Calibration
Measurement
Clicking this button starts the calibration process for all selected channels.
At the completion of the calibration, the measured mV and the Transducer
mV/EU value for the applicable channels are automatically updated.