SECTION 11—CONTROLLER
Part No 1122167
83
Zoom 300 and Zoom 400
Speed Lever in Neutral Menu
This menu is used to calibrate the throttle potentiometer for neutral such that when the
throttle is at rest (in neutral), the controller reads a voltage that translates to zero speed.
The default voltage for zero speed is 2.3 VDC (±0.2).
1. Select the Speed Lever in Neutral menu option.
2. Loosen the two (2) set screws in the throttle assembly.
3. Rotate the throttle shaft slowly until the buzzer sounds.
NOTE: If the controller version is 2.46 or higher, the Status LED will also glow.
4. Press ENTER to confirm the value when finished.
Controller Version Menu
This menu is used to display the current version of the Controller software.