SPECIFICATIONS
STANDARD SPECIFICATIONS
Electrical Requirements:
120 VAC +10% (132 VAC/108 VAC), 60 Hz
Rated Current Input:
4.0 A (ALL Models)
Sound Level:
52 dB Average (3LX)
53 dB Average (5LX, 5LXO2, 5LXO2T, 6LXO2, 6LXO2T)
Operating Altitude:
Up to 6,000 ft. (1828 meters) above sea level without degradation of
performance. From 6,000 ft. (1828 meters) to 13,129 ft. (4000 meters) below
90% efficiency.
Ambient Temperature Range:
50
O
F - 95
O
F (10
O
- 35
O
C)
S
P
E
C
I
F
I
C
A
T
I
O
N
S
SPECIFICATIONS
* Stated performance maximum achieved after 30 min. of run time.
Maximum Outlet Pressure:
(5LX, 5LXO2, 5LXO2T, 6LXO2, 6LXO2T)
(3LX)
5 psi + 0.5 psi.
5 psi + 0.5 psi.
(34.5 kPa + 3.45 kPa.)
(34.5 kPa + 3.45 kPa.)
SensO2 Alarm Thresholds:
* Factory Preset at 73%
Indicators
O2 Purity
Over 85% (+ 2%)
73% (+ 3%) to 85% (+2%)
Below 73% (+ 3%)
Internal Switch
*Set at 73%
GREEN Indicator Light
YELLOW Indicator Light
RED Indicator Light -
Continuous Audible
Alarm Sieve GARD
TM
Compressor Shutdown
Internal Switch
Set at 85%
GREEN Indicator Light
RED Indicator Light -
Continuous Audible
Alarm Sieve GARD
TM
Compressor Shutdown
N/A
*Oxygen Output Concentration
Levels:
ALL MODELS
95.6% to 87% at all flowrates.
If no internal selector switch, 73%
operation ONLY.
1/2 - 5 L/min. (5LX, 5LXO2, 5LXO2T, 6LXO2, 6LXO2T) 1/2 - 3 L/min. 3LX)
For flow rates less than 1/2 L/min., use the Invacare Pediatric Flowmeter
PF16.
460W (6LXO2, 6LXO2T) 350W/400W (3LX)
400W (5LX, 5LXO2, 5LXO2T)
35 psi + 3.5 psi (241 kPa + 24.1 kPa)
Flow Range:
Power Consumption:
Pressure Relief Mechanism
Operational at: