PROCEDURE 15 - REPLACING HOUR METER (FIGURE 1)
CAUTION
TO PREVENT ELECTRICAL SHOCK, ALWAYS DISCONNECT CONCENTRATOR FROM
ELECTRICAL OUTLET BEFORE SERVICING.
HOUR METER
PROCEDURE 15
H
O
U
R
M
E
T
E
R
1. Unplug unit.
2. Remove cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
3. Remove control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
4. Remove and code the two (2) wires connected to back of hour meter.
5. Spread retaining clips on hour meter housing that secure hour meter to control panel.
6. Remove hour meter by pushing meter through front of control panel.
7. Install new hour meter reversing STEPS 4-6.
8. Reinstall control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
9. Reinstall cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
Retaining Clips
(Hour Meter Housing)
Hour Meter
Hour Meter
Connectors
Hour Meter Housing
Back of Control
Panel
FIGURE 1 - REPLACING HOUR METER