52
PROCEDURE 14 - REPLACING FLOWMETER (FIGURE 1)
CAUTION
TO PREVENT ELECTRICAL SHOCK, ALWAYS DISCONNECT CONCENTRATOR FROM
ELECTRICAL OUTLET BEFORE SERVICING.
F
L
O
W
M
E
T
E
R
FIGURE 1 - REPLACING FLOWMETER
Palnuts
Flowmeter
1. Unplug unit.
2. Remove cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
3. Remove control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
4. Remove tygon tubing from the top and bottom fittings on back of flowmeter.
5. Remove palnuts from rear of flowmeter.
6. Remove flowmeter from front of control panel.
7. Install new flowmeter reversing STEPS 4-5.
8. Reinstall control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
9. Run unit and check for leaks where tygon tubing engages flowmeter fittings (Refer to PROCEDURE
19 - LEAK TEST).
10. Reinstall cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
PROCEDURE 14
FLOWMETER
Flowmeter Fittings
(Back of Flowmeter)
Control Panel