TROUBLESHOOTING GUIDE
T
R
O
U
B
L
E
S
H
O
O
T
I
N
G
UNIT NOT OPERATING
ALARM: OFF
Internal Status Indicators:
RED: Off
GREEN: Off
CONTROL PANEL INDICATORS:
RED: Off
YELLOW: Off
GREEN: Off
Purity Switch setting at 73% or 85%.
TROUBLESHOOTING GUIDE
NOTE: REFER TO THE REPAIR/REPLACEMENT GUIDE IN THIS MANUAL FOR PROPER
REPLACEMENT OR ADJUSTMENT PROCEDURES
1. Replace P.C. Board. (PROCE-
DURE 11). Unit may need
retiming after board replacement.
(PROCEDURE 18).
2. Reattach horn connector to
harness.
2. Horn.
Disconnected.
1. Purity Switch.
Defective.
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
1. Transformer Assembly.
a. Molex connector disconnected
from P.C. Board.
b. Faulty transformer assembly.
UNIT NOT OPERATING,
Internal Status Indicators:
RED: Three (3) Flashes
GREEN: One (1) Flash
(Integrated Circuit Board)
Control Panel Indicators:
RED: Off
YELLOW: Flashing
GREEN: On
UNIT NOT OPERATING,
Internal Status Indicators:
RED: Off
GREEN: Off
Control Panel Indicators:
RED: Off
YELLOW: Off
GREEN: Off
Power Switch ON. Continuous
audible alarm.
1. Internal repairs required.
1a.Reattach connector.
1b.Replace transformer assembly
(PROCEDURE 12).
1. Replace SensO2 circuit board
(PROCEDURE 11).