TROUBLESHOOTING GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
T
R
O
U
B
L
E
S
H
O
O
T
I
N
G
TROUBLESHOOTING GUIDE
LOW CONCENTRATION:
NOTE: Check for O2 purity
using a calibrated Oxygen Ana-
lyzer at Test Point 1 (OXYGEN
OUTLET) of the concentrator.
Internal Status Indicators:
RED: On GREEN: Off
(
LX Control Board)
RED: Two (2) Flash
GREEN : One (1) Flash
73% Shutdown
(Integrated Sensor Board)
RED: Two (2) Flash
GREEN : Two (2) Flashes
85% Shutdown
(Integrated Sensor Board)
Control Panel Indicators:
SensO2 ONLY:
RED: On
YELLOW: Off
GREEN: Off
For SensO2 units, the Red indica-
tor will signal extremely low purity
and will be accompanied by a
continuous audible alarm and a
system shutdown. Repairs are
required.
Internal Status Indicators:
RED: Off GREEN: Off
(
LX Control or Integrated Sensor
Board)
Control Panel Indicators:
SensO2 ONLY:
RED: Off
YELLOW: On
GREEN: Off
A Yellow indicator signals low
purity, but within specification for the
user. Repairs are required.
NOTE: REFER TO THE REPAIR/REPLACEMENT GUIDE IN THIS MANUAL FOR PROPER
REPLACEMENT OR ADJUSTMENT PROCEDURES
1. Clean or Replace (PROCEDURE
2).
2. Replace (PROCEDURE 2).
1. Cabinet Filters Dirty.
3. Compressor.
a. Defective.
b. Faulty Capacitor.
c. Bad Motor Windings.
d. Worn Seals.
e. Bad Bearings.
f. Leak at fittings or tubing.
3a.Replace Compressor (PROCE-
DURE 7).
3b.Replace Capacitor (PROCE-
DURE 7).
3c.Replace Compressor (PROCE-
DURE 7).
3d.Rebuild Top End of Compres-
sor (PROCEDURE 2).
3e.Replace Compressor (PROCE-
DURE 7).
3f. Replace Fittings or Tubing.
2. Inlet Filter Dirty.
4. Heat Exchanger.
a. Leak at tubing or body cham-
ber.
4a.Replace or Retighten.
4b.Replace or retighten tubing.
Replace Heat Exchanger (PRO-
CEDURE 8).
b. Inspect tubing and Heat
Exchanger.
7a.Leads to fan Disconnected.
Reconnect (PROCEDURE 10).
7b.Replace (PROCEDURE 10).
8. Replace (PROCEDURE 4).
9. Repair or replace.
10a.Check pressure at product tank.
Pressure should rise to 20 p.s.i.
at shift point. If not, replace P.C.
Board (PROCEDURE 21).
10b.Inspect tubing at transducer and
at product tank. Replace tubing
ONLY if damage is evident.
(PROCEDURE 9).
b. Pressure transducer tubing
leaks.
g. Leaky or Defective Relief
Valve.
h. Insufficient voltage (outlet).
3g.Replace Relief Valve.
3h.DO NOT use extension cords.
5. Replace (PROCEDURE 6).
6. Replace (PROCEDURE 2).
5. Regulator Cracked or Leaking.
6. Exhaust Muffler dirty or plugged.
7. Fan
a. Not Operating. Unit overheat-
ing.
b. Faulty Fan.
8. Sieve Beds defective
9. Tubing kinked or blocked.
10.P.C. Board
a. Shifts at wrong pressures.