![Invacare 3LX Service Manual Download Page 50](http://html1.mh-extra.com/html/invacare/3lx/3lx_service-manual_2082378050.webp)
50
PROCEDURE 12 - REPLACING TRANSFORMER ASSEMBLY (FIGURE 1)
CAUTION
TO PREVENT ELECTRICAL SHOCK, ALWAYS DISCONNECT CONCENTRATOR FROM
ELECTRICAL OUTLET BEFORE SERVICING.
T
R
A
N
S
F
O
R
M
E
R
1. Unplug unit.
2. Remove cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
3. Remove control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
4. Locate the transformer assembly in the upper right corner on the back of control panel.
5. Unplug the transformer connector from the Control (bottom) P.C. Board.
6. Remove the two (2) keps nuts that secure transformer to the back of the control panel.
7. Reverse STEPS 5-6 to install NEW transformer.
8. Torque keps nuts to 13.5 + 1 in. lbs.
9. Reinstall control panel (Refer to PROCEDURE 9 - REMOVING CONTROL PANEL).
10. Reinstall cabinet (Refer to PROCEDURE 1 - REMOVING CABINET).
Transformer
Connector
Keps Nut
Transformer Assembly
Keps Nut
Control (Bottom) P.C. Board
(Typical)
FIGURE 1 - REPLACING TRANSFORMER ASSEMBLY
PROCEDURE 12
TRANSFORMER
Transformer P.C. Board Connector
(All Models)
Control Panel