background image

.

6. Accessory I/O

On the middle right side of the instrument rear panel is

the 24 pin card edge connector, marked

‘ACCESSO

RY

INPUT’

 or ACESSORY INPUT OUTPUT (on newer

models). This connector permits access to circuitry for test
purposes, as well as for special user functions. The two
most important functions are remote control and auxiliary
power, as described below.

6.1 Remote Control

Remote operation for continuous wave (D.C)

measurement s requires only the RS232C port, with the
function set to the auto-range mode. Data will be
transmitted when requested via the RQS (ReQuest to Send)
line on that interface.

Flash measurement s require more control over, and

coordination with the instrument, through the

‘ACCESSO

RY

POR

T’

. Either external switch control or standard computer

logic will activate the four functions equivalent to the front
panel push buttons. The (inverted) FLAS output, goes low
once the ILT1700 is ready to integrate. This logic level can
be used to trigger the flash lamp, or

‘handshake’

with a

computer device to complete a two way protocol. The
following is a list of user inputs:

Pin#

Mnemonic

True Level Description

10

/ZERO

Low

Zero (not)

L

/SWDN

Low

D.C. (not)

J

/INTG

Low

Integrate (not)

K

/HOLD

Low

Hold (not)

9

/FLAS

Low

Flash (not): output

If mechanical activation is required, a switch closure

between any of the above listed inputs, with respect to
ground, will activate that function. It is only necessary to

keep the switch closed for 50 milliseconds to assure
sampling response from the internal computer. Contact
bounce less than 1 millisecond is allowed for reliable
operation.

Digital logic, or the output from a computer user port,

can be used if it is TTL compatible or if the drive can

‘sink’

at least 1 milliamp, to a voltage level below 0.8 volts, and
allow the level to rise above 3.5 volts (essentially no
sourcing necessary) upon release. The low state is the

‘true

level

in each case. The description, on the previous chart,

shows which function key the low state will equate to.

When the ILT1700 is ready to integrate, it will pull pin 9

to a voltage level less than 0.8 volts (assuming the current
load is less than 1.3 mA). This occurs in less than 8
milliseconds after /INTG is pulled low. That line can be
used to activate a flash lamp for fully automatic operation.

A fully automatic system would pull

‘/ZERO’

low as

part of the initializatio n routine to cancel out any ambient
conditions. Some time later, when a test lamp was in
position,

‘/INTG’

would be pulled low for more than 50

milliseconds, or until

‘/FLASH’

was sensed. Three seconds

after that

‘/HOLD’

would be pulled low for 50 mS, and RQS

would be sent to the RS232C port until serial data begins to

be received. The data would be processed, and a new lamp

would be inserted in the test station, and

‘/SWDN’

would be

pulled low to zero out the integration register.

‘/INTG’

would be pulled low again to start the next cycle, and so on.
The

‘AU

TO-RANGE

feature should be on to handle any

magnitude of light source, and the

‘5

V BIAS

generally

should be on to increase the detector speed and to handle the
large instantaneous charge from the flash.

6.2 Test Pins

There are accessory pins that are used during the

system test at the factory, but a user normally will not be
concerned with any of these except possibly for pin 11
(VRAM), which tests the

‘kee

p alive

voltage for the

continuous memory. This should be greater than 2 volts
when measured with a high input impedance volt meter, with
respect to ground (found on pins A, or 1). Pin 3 is the +5
volt logic power supply voltage, which can be used for
external logic as long as it draws less than 50 milliamps.

6.3 DC Power

When the internal batteries are used, the instrument

automatically turns itself off after about 6 minutes. For
field testing, continuous measurement s may be desired or it
may be beneficial to recharge the internal Nickel-Cadmium
batteries from a mobile power source. In either case the
Auxiliary input can be used as found on pin C. This
auxiliary D.C. (direct current) power must have a potential
b8 and +15 volts with respect to the ground pins(A

or 1). This range was chosen to permit use of normal

mobile power. We offer an accessory called the A401,

which provides a cigarette lighter plug, cable, and accessory
connector for such mobile use. If mobile power is not
available, the instrument may be operated from any 12 volt
battery capable of supplying 380 milliamps for the duration
of the experiment.

6.4 Analog Output

An analog output pin is available on t he accessory
connector.  It is brought out on pin 8, which uses any of the
ground pins 1, A or N, as the signal return path.  This signal
called RECO.

7. Precautions

The ILT1700 has been designed to minimize problems

due to improper operation of the instrument. From our 40
years of manufacturing instruments, we have found a few
abuses that can cause trouble, as follows:

19

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Summary of Contents for ILT1700

Page 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Page 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Page 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Page 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Page 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Page 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Page 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Page 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Page 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Page 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Page 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Page 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Page 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Page 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Page 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Page 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Page 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Page 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Page 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Page 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Page 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Page 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Page 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Page 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Page 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Page 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Page 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Page 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Page 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Page 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Page 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Page 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Page 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Reviews: