International Light Technologies ILT1700 Instruction Manual Download Page 27

26

the light is picked up from a small spot in the image plane.
This is very nice for measuring the brightness of an
illuminated segment of an alpha numeric display, or for
measuring the dot brightness on the face of a CRT.
Unfortunatel y these systems are very expensive. Another
alternative is to use simple lenses to image a small portion
of a test field, onto an aperture which has a detector behind
it. This is very effective especially if the source is a
repetitive configuratio n in a production situation. A custom
set up can be made to specifically measure that one small
emitting surface. Our Radiance accessory has a 1.5 degree
field of view, with the objective lens being about one inch in
diameter. This is effective as long as the target is larger than
one inch. If you must back up from it, the target requirement
gets bigger by the distance times 2*(tan 1.25), plus some
margin for error. If you were a foot away, the target should
be at least 2 inches in diameter.

8.9.4 Uniform Receiver Sensitivity

As mentioned in section 8.3, the best receiver for

uniformity is the input port of an integrating sphere. The
uniformity we now are talking about is that which is
measured perpendicular to the optical axis, over the input
surface. For Flux measurements it is necessary to have this
uniformity, so that small errors in centering the beam do not
contribute to much of a change in the measured reading.
Our narrow beam adapter (HNK15) attachment, is designed
to accept a few millimeters of axial misalignment without
appreciable changes in the reading. This is necessary to
allow for non critical positioning of laser beam
measurements.

8.10 Temporal Response

This refers to the light time response. There are many

factors that should be considered when measuring fast light
pulses, including the need for instantaneous information, or
whether dose information is adequate. To obtain
instantaneous data, the entire electrical system, including the
final oscilloscope, must be properly designed and matched.
This includes the connecting cable, characteristic
impedances, and matching amplifiers. In most cases it is
necessary to put a preamplifier in the detector head to match

the high impedance of the detector to the coaxial cable.

When you do this, the dynamic range is limited to a few

decades at best, and auto ranging is not practical. For these
reasons we have chosen to concentrate on dose measuring,
which generally matches the ultimate goal of the light pulse,
and is compatible with the same detectors used for c.w.
measurements. When we say

‘ultimat

e

goal’

, we mean that

the energy in a pulse is generally the factor that determines
its effectiveness to perform work, and therefore is the best
figure of merit when making a measurement. Capacitance is
one of the properties of a detector that is very detrimental to
fast instantaneous measurements, but does not produce an
error when integrating, since it just tends to store the charge
for subsequent removal. By designing charge measuring
electronics into the instrument, we can rely on charge

storage in both the detector and in a capacitor on the front
end of the system. By measuring with a 5 volt reverse bias
on the detector, we now eliminate junction saturation due to
internal series resistance voltage drop in the detector, plus
we store the charge in the capacitance of the detector
(approx 4 nanofarads for SED033), the coaxial cable
(approx 0.23 nanofarads), and the 390 nanofarad capacitor
located on the front end of the amplifier. At 5 volts bias, the
system can temporarily store more than 1 microcoulomb of
charge, which is quantized in one measurement cycle (half
of a second). The lowest charge of 10 picocoulomb, will
still have a 5% resolution, providing a pulse dynamic range
of five decades. There are other limitations, such as the
maximum current output from the detector, and the intrinsic
speed of the detector itself which will be covered separately
below.

8.10.1 Low Duty Cycle (fast pulse)

The ILT1700 can measure a single fast pulse as long as

the pulse rise time is generally greater than 5  microsecond,
and the peak detector current is less than 2.0 milliamps.
These limitations refer to the SED033 detector, which
generally gives the best performance for flash
measurements. The vacuum photodiodes are very fast
devices, but the peak output current is about 1.0
microamperes, which is a limitation to high speed,
high magnitude light sources. (attenuators can be used to
decrease sensor sensitivity) If one wishes to measure a
continuous stream of pulses, the ILT1700 has a unique
capability of averaging the reading, even if the duty cycle is
extremely low. This of course is also limited by the
boundaries of the peak current, and intrinsic detector speed.
Another method for measuring low duty cycle signals, is to
integrate while counting the number of flashes. Then divide
the total integral by the number of flashes to get the average
charge in one flash. Still another technique is to integrate
for a known period of time, and then divide by the number
of seconds to get the average reading/sec. Data is
updated exactly every half second.

8.10.2 High Peak Amplitude

If you have a very bright C.W. source that  would produce

more than 2 milliamps from the detector, the instrument will

read “HI” on the display.  If the peak current is the result of

a

high amplitude flash, the limit usually is due to the detector
saturation.  The most often used silicon detector (SED033) can
operate up to 2.0 peak milliamps before saturation, while the
vacuum photodiodes (SED240 or SED220), have a peak less
than 1.0 microamperes. In flash mode, there system may not

read “HI”

because the measurement integral does not detect the

peak. ILT offers attenuation filters (QNDS1, QNDS2, QNDS3)
for attenuation by a factor of 10, 100, and 100 respectively. To
test, simply add the QNDS fil ter, without changing the
calibration factor and the output should drop by the same
attenuation rate as that of the filter.(ie QNDS1 x 10)

You can easily determine the current levels:.

 1. enter a calibration factor of 1, and select that factor, or
2. multiply the readings taken by the calibration factor being
used. Both provide the detector current in amperes.

For further applications assistance, check out  our
complimentary handbook on our website.

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Summary of Contents for ILT1700

Page 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Page 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Page 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Page 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Page 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Page 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Page 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Page 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Page 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Page 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Page 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Page 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Page 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Page 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Page 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Page 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Page 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Page 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Page 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Page 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Page 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Page 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Page 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Page 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Page 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Page 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Page 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Page 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Page 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Page 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Page 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Page 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Page 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Reviews: