background image

3.2.2 Range Selection

Most users will find the auto range mode the best choice

when using the Radiometer. This is the initializatio n default
mode which will display any magnitude automatically, as
long as a proper previous zero had been performed (see the
previous section) and the correct sensitivity factor is
programmed and selected (see the next section). There are
two other range choices, that may apply for special
circumstances. I f many repetitive measurements are to be
made, it is often desirable to turn the auto range mode off,
which essentially freeze s the range exponent, so the user is
alerted to a large change in signal by the fact that the
mantissa will either show several zero digits, for a low
reading, or will read

‘HI’, t

o indicate an upward range

change. In scientific notation, yo u could experience a
tenfold change in the light level, which might go unnoticed
because the exponent is the only obvious digit that changed.
By turning the automatic feature off, it becomes more
obvious that a significant change has occurred. Another
reason to turn the auto range off is to read a zero that occurs
quickly as described under

‘zeroing i

n section 3.2.1 above.

The third range choice is

‘percent mode’

, and is described in

section 3.2.4.

3.2.3 Programming

The Radiometer performs subtraction, as mentioned in

the zeroing section (3.2.1). It also calculates ratios as
described in the next section (3.2.4), but for most
measurement s it is required to scale the answers by a
sensitivity factor in order to read directly in your desired
optical units. This scaling is done internally by dividing the
electrical current coming from the input device by one of ten
(10) sensitivity factors, and then displaying the result in
floating point arithmetic, on the front panel. These answers
can range from 0.00x10

-19

to 9.99x10

+19

.

3.2.3.1 Current Measurements - If you want to read the

electrical current, you may select a factor of 1.00x10

+0

(which is 1). The answer will be displayed in amperes.
These are good units for any sensor that is not calibrated,
because a subsequent calibration would permit you to
reconstruct the absolute data. The best policy is to obtain a
calibration sensitivity factor for the detector assembly at the
time of purchase, so that direct readings can be obtained
immediately.

3.2.3.2 Entering Factory Calibrated Factors - The

factory calibration provides a sensitivity factor that appears
on the calibration certificate in units with amperes (A) in the
numerator and the desired readout units in the denominator.
When the instrument divides the current (amperes) by this
factor, the ampere term drops out leaving the final desired
optical units on the display.

You may store ten factors (in registers 0 to 9) for future

selection. Each factor is designated by a number in the factor
select window. (See 3.2.3.3 below) We suggest that you
store 1.000x10

+0

as one of  the entries, for reading in units of

current(amperes)
To enter a factor, turn the instrument on, toggle

the ‘

 DATA

DISPLAY

button to

FACTOR SHIFT

, then increment the

FACTO

R’

to read the register you wish to change (0 to 9).

Now use the three black buttons

marked ‘MSD’, ’LSD’

,

AND

‘EXP’

to enter the number in scientific

notation, just as it appears on the calibration certificate. By
holding a button in, the display will begin to roll faster, so
you can rapidly increment to a number some distance away.
As you approach the desired number, pulse the button one
step at a time

3.2.3.3 Sensitivity Factor Selection (0-9) - The last

window is the

FACTOR SHIFT

readout. It shows the

user which of 10 stored factors has been selected. Each
detector/filter/optic combination has a unique calibration factor
therefore one detector can require several factors for
different combinations or readout units . Also at different
wavelengths there may be additional sensitivity factors, so
the wavelength would determine the

FACTOR SHIFT

number used. This is changed by switching to the FACTOR

mode on the DISPLAY switch, and then incrementing the

FACTOR SHIFT

button to arrive at the desired factor

number.

3.2.3.4 Self Calibration Technique - These factors may

also be used for customizing readouts in relative units that
pertain to your particular system. For example; an optical
throughput, in a hypothetical system, may be adjusted at the
factory to a particular magnitude, such as 150 (1.500e

+2

on

display). By selecting the right factor, you can make the
instrument also read 150 for that desired magnitude. The
ILT1700 can then be used for a subsequent quality control
check to adjust production units to have the same nominal
output. Another relative application could be to obtain a
precise correlation between departments. A standard source
would be used as a reference so all measurement systems
would be set to read the same from the same source. This
type of

‘i

n house

calibration can provide an accuracy better

than one percent, without having to refer back to the
reference source very often, especially if using a silicon
detectors which are generally more stable than standard
sources.

Self Calibrating Procedure: (ZERO METER FIRST)
A. Enter the ultimate display reading into one of the 10
factor registers, by pressing the DATA DISPLAY button
to

FACTOR SHIFT

mode, and by using the MSD and

LSD buttons to enter the number (in scientific notation).
B. Switch back to the

‘D

ATA

mode and take a reading.

Record this on paper for the next step.

C. Enter the reading from step

‘B’

, into one of the 10

registers as you did in step

A

. You can use the same

register, the last number is no longer needed.
D. Now switch back to the

‘D

ATA

mode again and read

the same optical signal. It should display the number
you were trying to calibrate to in the first place.

3.2.4 Percent Mode

There are two very common optical uses for the percent

mode, transmissio n and reflectance. Both measurements
require the use of a stable light source. By stable, I mean
that it will remain constant during the measurement
procedure. Modest accuracy can be achieved with a line
powered lamp, but for good long term repeatability, we
recommend an electronic regulator.  (The visible output
from an incandescent lamp changes 3% for each 1% change
in voltage). The first step is to set up baffles so the light
travels directly from the lamp

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Summary of Contents for ILT1700

Page 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Page 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Page 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Page 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Page 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Page 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Page 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Page 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Page 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Page 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Page 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Page 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Page 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Page 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Page 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Page 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Page 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Page 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Page 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Page 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Page 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Page 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Page 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Page 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Page 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Page 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Page 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Page 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Page 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Page 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Page 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Page 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Page 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Reviews: