background image

3.3.5 Readout (Scientific Notation)

The display reads the total dose in scientific notation

(see section 3.2.5 for an explanation). When integrating, the
readout is a summation of light over a period of time. The
units of readout become the product of the Illuminance or
Irradiance times time. As an example, suppose you were
integrating to determine the exposure for UV curing. The
detector would be reading instantaneous irradiance in units

of Watts per Square centimeter (W/cm

2

), but the exposure

would be in

Watt seconds per square centimeter

(W*sec/

cm

2

), and since a Watt second is equal to a Joule it would be

in Joules per square centimeter(J/cm

2

). If you are dealing

with small doses, the answer may be expressed in millijoules
per square centimeter(mJ/c m

2

), by multiplying the answer

by 1000, and similarly in microjoules per square centimeter
(µ J/cm

2

) by multiplying by 1,000,000. For example if the

instrument displays

‘1.234

e

-2

J/cm

2

, you can multiply by

1000 to get 12.34 mJ/cm

2

as your answer.

3.3.6 Darkroom Integration

When integrating in the dark, you must be able to easily find
two buttons.  The first is the

‘INT’ which can easily be found by

placing the palm of your hand on the rig ht side of the instrument
and feeling over for the second button from the right with your
thumb.  In a

similar manner, you will locate the ‘HOLD’ button

to stop the integration.  This is the easiest to find since it is the
first button in from the right si

de.  Once ‘HOLD’ has been

pressed the lights may be turned on to make the readout
without changing the reading.  It is also possible to operate
the instrument with a remote cable from another room, and
to print the answers on a small printer, or by sending the data to
a computer over the RS232C  or USB interface (see sections 4.
& 6. for more on this).

3.4 Bias Selection

As mentioned in section 2.1 (front panel controls), there

is a bias button labeled

‘5

V

BIAS’

. This raises the input

node of the input amplifier up to +5 volts which, in turn,
reverse biases a silicon detector with 5 volts. If it is a
vacuum photodiode, it adds another 5 volt bias to the 9 volts
that is already applied to the cathode of the phototube,
making the total reverse bias of 14 volts. See the following
sections for more details.

3.4.1 Vacuum Photodiode Bias

Vacuum Photodiodes always require a bias voltage to

operate properly. All of the International Light Vacuum
Photodiodes will operate properly with a bias between 9 and
75 volts. We recommend operating them at the 14 volts
provided with the bias ON, so the readings correlate with
our calibration lab, and with previous instruments that
applied 12 and 15 volts. Another advantage to the larger
voltage is the improvement in peak current capability, for
maximizing the dynamic range of flash measurements. The
difference in D.C. readings between 9 and 15 volts bias, is
only about 1%, but to gain that extra accuracy, use it with
the

‘5

V BIAS

light on.

14

3.4.2 Flash Bias

Detector bias has the most significant effect on system

performance in flash measurements. Semiconducto r devices
gain in three ways when placed in a reverse biased
condition, as opposed to the photovoltaic mode (zero bias).
The first improvement is to reduce the junction capacitance,
which makes the instantaneous photo current get to the
photometer faster, rather than being delayed as a junction
charge. The second improvement is the elimination of
junction saturation. This happens because the instantaneous
photocurrent produces an I*R drop across the top
transparent electrode, which in turn allows the junction
voltage to become forward biased, causing junction
saturation. The third advantage has to do with the method
used to measure the charge from a flash. In order to make
the radiometer operate at speeds faster than the clock time of
the computer, it is necessary to temporarily store the flash
charge, so the charge digitizing circuitry can withdraw this
charge and measure it. This temporary storage can handle
2.00e-6 coulombs, so the instrument maintains 6 decades of
dynamic range even down to nanosecond speeds. Below 100
microseconds the limitation is due to the peak current of the
detector, not the instrument. Our SED033 is designed to
handle more than 2.0 milliamps, to still provide 3 decades of
useful range at 1 microsecond. By combining this basic
capability along with a neutral density attenuator, one can
make fast measurements, spanning four or five decades.
Anything below one microsecond is not recommended, since
a multitude of detector speed limitations, including
impedance matching etc., cause inaccuracy.

3.4.3 Low Level Bias

When making low light level measurements, it is

important to minimize the detector leakage current. With a
semiconductor, this can be done by using it in the
photovoltaic mode, with NO BIAS on it at all. This also
minimizes the temperature effects on the detector due to
changes in the internal shunt resistance. Vacuum
Photodiodes must always have a bias on them, so they
should be operated with the

5 V BIAS

light ON. As part of

the detector design, we have minimized detector leakage
through the coaxial cable by keeping the current carrying
lead at the same potential as the shield lead, even in the
reverse bias mode. Vacuum Photodiodes have a good low
light level attribute, the absence of 1/f noise. This low
frequency (thermal) noise, can be a serious problem for D.C.
measurement s when using semiconductors. The vacuum
photodiode noise advantage tends to compensate for the
higher semiconductor responsivity, making them continue to
be a very effective transducer, especially for the short wave
length measurement s in the UV, and for their ability to reject
the longer unwanted wavelengths.

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Summary of Contents for ILT1700

Page 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Page 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Page 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Page 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Page 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Page 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Page 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Page 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Page 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Page 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Page 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Page 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Page 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Page 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Page 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Page 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Page 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Page 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Page 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Page 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Page 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Page 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Page 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Page 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Page 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Page 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Page 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Page 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Page 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Page 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Page 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Page 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Page 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Reviews: