30

Data Displays: Above the graph are three additi onal displays.  The Left screen is the irradiance display, the

center screen is the percentage display and the right is the integrate display.  These displays will change

values based on the features you select using the DC, SET 100%, and INT buttons.

If you need to do a software zero, cover the detector connected to the ILT1700 with the black cap and

press the Zero button

. The LED indicator will turn  green until complete and the Zero Level

indicator should read

.  Assure the meter was already zero-ed prior to using the program.  Once

you have zero-ed, the program should show readings as displayed on the ILT1700 meter.

The data can be saved at any point in a tab -delimited file by pressing the Save button

.

Carefully  read  the  pop  up  messages  and   follow  the  instructions.    To  stop

  the  program

without  saving  the  data,  press  the Stop  button.    To  measure  in  percentage,  click  the Set  100%  button

 to set the baseline 100% level.  All future readings will be in percentage as a comparison to the

100%  baseline  reading.    To  measure  accumulated  dosage  or  total  exposure,  press  the Integrate  button

.  The program will continue to integrate and show the total dosage in the integration window

until  you press Save or Stop.  To exit or stop the program, always use the Stop or Save buttons as they

assure that the COM port is properly closed.  Try not to use the Abort Execution button

 in the VI

toolbar.

E. TROUBLE SHOOTING

An

error  message  in  “VISA  Read  in  ILT1700.VI”will  appear  in  case  that  a  wrong  COM  Port

is

selected  or  if  the  VCP  driver  is  not  properly  working.    To  exit  the  message,  click  Stop  and  select  the
correct Com Port.  Use Control Panel \System\Hardware\Device Manager\Ports in the Start Menu to locate
the proper COM port or to check the proper worki ng of the VCP driver.

F. HARDWARE FEATUR ES

USB 2.0
Windows 2000, XP and VISTA
Bits per second: 4800
Data bits: 8
Parity: none
Stop bits: 1
Flow control: none

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Summary of Contents for ILT1700

Page 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Page 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Page 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Page 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Page 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Page 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Page 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Page 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Page 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Page 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Page 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Page 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Page 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Page 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Page 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Page 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Page 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Page 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Page 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Page 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Page 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Page 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Page 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Page 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Page 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Page 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Page 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Page 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Page 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Page 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Page 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Page 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Page 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Reviews: