background image

25

8.8.1 Specular Reflectance

All reflectance measurement s require some special

fixtures. In this case you must have a detector holder that
can be swept in an angular arc of 90 degrees about a
rotational point which has a holder for the flat sample
reflector. A stable light source is directed across the top of
this rotational point to some distance (x) which hits the
properly filtered detector. The button called

‘SE

T 100%

is

pressed to establish the unattenuated condition for reference.
The reflector is then placed directly over this rotational
point, at an angle of 45 degrees to the source. The detector
is rotated, also about this same point, for an angle of 90
degrees or until a peak output is found by watching the
meter readings. It is important that the distance (x) is still
the same as it was before moving the detector. The sample
surface must be flat in order to reproduce the same beam
divergence as present without any reflector present. There
are many variations to this method, depending upon the
ultimate use of the reflector. Obviously if the reflector is a
curved surface this would not work. It may be necessary to
use a setup similar to the diffuse measurement in an
integrating sphere, or use a goniophotomete r to integrate the
output over the entire divergent reflectance angle. These
methods are beyond the scope of this manual however.

8.8.2 Diffuse Reflectance

In the previous paragraph, we mentioned two techniques

for measuring diffuse reflectance, namely by use of an
integrating sphere or by integrating the total reflectance
using a goniometer. Both methods require very special
equipment, and will be lightly discussed here, since it would
be impossible to do the subject justice in a document of this
nature. In the first case, a collimated light beam projects
through an integrating sphere, out an opposite port on the
other side. A detector, with the desired spectral response, is
placed in the sphere surface orthogonal to this beam, so as to
be blind to either of the other ports. The

‘ZERO’

button is

pressed with the sample port open. Then a white reflectance
standard is placed in the sample port, with a surface such as
barium sulphate, magnesium oxide etc. and the

‘SE

T

100%’

button is pressed. Now the standard is removed, and the
sample is replaced in the sample port. The display on the
ILT1700 then reads the diffuse reflectance relative to the
standard reflectance. If the standard had a reflectance of
98% then you must divide the answer by .98 to get the
absolute sample reflectance. This type of measurement does
not measure the specular component, since it is reflected
back out the beam entrance port and lost. Other angular
geometries are required to include the specular component.
The application dictates what is to be measured.

8.9 Spatial Response

Spatial response refers to the change in responsivity as a

function of translational displacement s (x or y), and with
angular displacements (pitch, yaw, and roll). Or the measure

of a detector’s relative sensiti

vity as a function of incident

wavelength.

8.9.1 Lambertian Response

Lambertian response is in reference to a particular

angular response proportional to the cosine. In other words
the cosine of the angle normal to the face of the detector is
one or 100%. As the angle goes off axis and becomes
parallel to the face of the detector, the reading goes to zero,
as does the cosine of the same angle (90 degrees). At 45
degrees the cosine is 0.707, which means that the detector
should read the rays with 70.7% of the value produced by
the same rays entering normal to the input device.

The reason this spatial response is necessary for

accurate measurements is that it matches the spatial response
of a perfect absorbing surface. Since Irradiance and
Illuminance, are measurements of light falling on a surface,
the cosine is compatible with these measurements. An
analogy of the perfect absorber might be considered as being

a small hole in a piece of sheet metal, placed over a well.

All the light that goes in that hole will be absorbed by the

deep well hole underneath. None will get reflected back up
out of the same hole. If we analyze the effect of a change of
angle, such as the sun moving from high noon to sunset, we
will see that less light can make it into the hole at sunset,
because the effective area of the hole is smaller as you view
it from an oblique angle. This reduction in area is directly
proportional to the cosine of the angle normal to this
surface. On polar plotting paper, the cosine makes a circle,

which is convenient when comparing the ideal response with

that of an actual plot.

8.9.2 Field Baffle

There are times when you should restrict the field of

view to delete oblique angles. In a lab environment, you
may be working with a light source on an optical bench.
The only light of interest is from that source, yet light
bounces off the people in the room and back to the detector,
creating errors in the readings. This means that you are
better off to restrict the field of view if you know there are
no sources to be measured, at the oblique angles. This can
be done with external baffles, or with our accessory hood
(H). Baffles can be made from sheet metal cut to form a
sharp edged hole in the middle. A square hole is actually
better than a round hole, since it is less likely to create
reflections in a multi-baffle array. Also, black velvet is
excellent for dividing off test areas from the rest of the
rooms lighting. If it is necessary to have light travel down a
tube, you can thread the inside of the tube to reduce the wall
reflections.

When making luminance or Radiance measurements, it

is absolutely necessary to restrict the field of view to one
that

‘sees’

only an intended test area of a reflecting surface,

or rear lit surface. Baffles can be used to implement this
kind of measurement without resorting to expensive optics.

8.9.3 Narrow Angle (Luminance/Radiance)

As just mentioned, there is a requirement for a narrow

field of view when making Luminance and Radiance
measurement s (see section 8.5). This can be accomplished
with lenses as in our Radiance barrel (R) accessory. In some
applications it is accomplished by using a telescope where

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Summary of Contents for ILT1700

Page 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Page 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Page 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Page 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Page 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Page 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Page 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Page 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Page 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Page 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Page 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Page 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Page 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Page 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Page 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Page 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Page 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Page 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Page 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Page 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Page 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Page 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Page 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Page 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Page 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Page 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Page 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Page 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Page 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Page 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Page 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Page 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Page 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Page 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Reviews: