INSTRUCTION MANUAL
Sahara Moisture Transmitter
Reference IR-085
Rev 3
Page 7 of 16
01/10/2014
V. Specifications
Moisture Range
Curve B: -148
to 86
F (-100
to 30
C)
Curve A: -112° to 86°F (-80° to 30°C)
Operating Temperature
-40
to 140
F (-40
to 60
C)
Storage Temperature
158
F (70
C) maximum. The probe should be stored with the plastic cover and
desiccant packet threaded onto the probe. Store in a cool, dry environment.
Warm-up Time
Meets specified accuracy in 3 minutes.
Accuracy
±
2
C from -85
to 104
F (-65
to 40
C)
±
3
C from -148
to -87
F (-100
to -66
C) for Curve B
±
3
C from -112° to -87°F (-80° to -66°C) for Curve A
Repeatability
±
0.5
C from -85
to 104
F (-65
to 40
C) dew/frost point
±
1.0
C from -148
to -87
F (-100
to -66
C) for Curve B
±
1.0
C from -112° to -87°F (-80° to -66°C) for Curve A
Response Time
Less than 5 seconds for 63% of a step change in moisture content in either wet-up or
dry-down cycle.
Summary of Contents for HC-103
Page 2: ...USER MANUAL Series HC Reference IR 222 Rev 3 Page 2 10 17 2013 ...
Page 31: ...USER MANUAL Series HC Reference IR 222 Rev 3 Page 31 10 17 2013 SECTION 4 MAINTENANCE ...
Page 50: ......
Page 52: ......
Page 54: ......
Page 56: ......
Page 69: ...SAHARA MOISTURE TRANSMITTER Instruction Manual ...
Page 85: ...USER MANUAL Series HC Reference IR 222 Rev 3 Page 46 10 17 2013 SECTION 8 CUT SHEETS ...
Page 94: ...Page 4 MODEL 1700TM DIMENSIONAL PRINT 050205 Model 1700 IG ...
Page 119: ...USER MANUAL Series HC Reference IR 222 Rev 3 Page 47 10 17 2013 SECTION 9 DRAWINGS ...
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 128: ......