Maintenance
99
1
Depending on the contamination the air filter should be cleaned or replaced as
necessary.
2
Push in new (changed) air filter until fully in contact with dead stop.
3
Close air filter fastener.
4
From the main menu go to
HISTORY
&
MAINTENANCE
/
CONFIRM
MAINTENANCE
and
confirm with the appropriate button on the right side of the display that the air filter
has been replaced / cleaned. The actual date and time as well as the actual
number of operating hours will be saved and a reminder to perform this
maintenance task again will be issued after another 10.000 operating hours.
Fig. 70
Dismounting of the air filter
Fig. 71
Inserting the air filter