Equipment Settings
53
Name
(Gas equivalent)
For displaying the leak rate in refrigerant equivalent leak rates go to the
EDIT
GAS
sub
menu and go to the
NAME
line item. Press the
EDIT
button and in the opening
GAS
SELECTION
submenu choose the refrigerant equivalent from the
GAS
LIBRARY
. Gas
parameters can be fully deleted by setting the gas name back to blank („<----->“).
Notice
If more than one set of parameter is set up (i.e. maximum two gases are
not named „<----->“) a list button will appear on the right side of the
measurement screen. When pressing the list button a short-cut to selecting
the different parameter sets will be provided.
Additional Helium
This submenu only is available when operating in gas equivalents. In this case the
Protec P3000 may display the Helium leak rate in addition to the refrigerant equiva-
lent leak rate in the main measurement screen. This feature may be selected or not
selected.
Default: not selected
Notice
The option “Additional He” is only available if at least one parameter set
(gas 1 … gas 4) is configured to helium.
The additional helium will be displayed in the unit of measurement of the
lowest gas number set to helium.
Fig. 41
Selecting additional Helium