How the Protec P3000 Works
31
In cases where the main unit is placed in a difficult or inconvenient to access area
the built-in PRO-Check reference leak can be removed from the main unit for easier
access and connected to the main unit via the Sub-D connector with a commonly
available extension cord. The PRO-Check reference leak can then be placed in an
area where it is convenient for the operator to be reached.
Notice
The PRO-Check reference leak is not part of the Protec P3000 shipment
and needs to be ordered as a separate part no. (see section
).
Notice
If you have not purchased the PRO-Check reference leak, warning 71 (“No
communication with test leak”) will be issued on first start-up. Please go to
SETTINGS
/
INTERFACES
/
PRO
-
CHECK
and set the PRO-Check to “
DISABLED
”
to prevent future warnings (see section
)
Notice
PRO-Check Warntime Expire Date (See Section 7.7.5).
Fig. 16
Connections for remote use
Pos. Description
Pos. Description
1
Connector at built-in test leak
2
Connector at Protec P3000 housing
1
2