Protec P3000 Messages
83
W4
24V of the OPTION socket
too low
Fuse F4 on the motherboard
has blown
Replace fuse F4 on motherboard*
W5
5V of the sniffer is too low
Fuse F5 on the motherboard
has blown
Replace fuse F5 on motherboard*
W6
5V of the internal Leak too
low
PRO-Check electronics
defective
Replace PRO-Check reference
leak or call nearest INFICON
service representative!
E7
-15V of the MC50 is too low
Motherboard defective
Call nearest INFICON service
representative!
E8
15V of the MC50 is too low
Motherboard defective
Call nearest INFICON service
representative!
E9
High Voltage error
Problem in sensor electronics
Call nearest INFICON service
representative!
W11
Wise current unstable
Current from Wise Technology
sensor did not stabilize within
20 min. after start-up!
Protec P3000 possibly not
being used for serveral days,
sensor current should stabilize
with more run-time.
Restart the Protec P3000.
If problem persists, call nearest
INFICON service representative!
E12
Wise Sensor not ignited
Current from Wise Technology
sensor too low for more than 10
min after power on
Restart the Protec P3000, if
problem persists, call nearest
INFICON service representative!
E13
Problem in Heater Control
Heater control for Wise
Technology sensor defective.
Call nearest INFICON service
representative!
W14
Discharge gone out
Sensitivity problem of Wise
Technology sensor
or
No ambient helium present
(e.g. sniffer purged with
nitrogen)
Warntime every 2h.
CAUTION
Make proof or calibration
Restart with sniffer line connected
to fresh air
If problem persists, call nearest
INFICON service repesentative!
W15
Replace sniffer tip filter
Filter in the sniffer tip is clogged Replace the filter in the sniffer tip!
W17
Fore pump service interval
expired!
> 10,000 operating hours since
last fore pump service
Replace membrane of diaphragm
pump!*
W18
Air filter service interval
expired!
> 10,000 operating hours since
last main air filter service
Clean or replace main air filter!
See Section
E19
No communication with
Wise ADC
Wise Technology sensor
defective or
CPU-board defective
Call nearest INFICON service
representative!
Error
no.
Plain text message
Possible Reason
Recommendation