BAUD
[Factory default = 19,200]
This sets the baud rate for the RS485 and the RS232 serial com-
munications. The baud rate can be set to various values through
the serial interface or via the front panel soft-keys. The parity can
only be changed through the serial interface command set.
When this occurs, the current setting will be shown in the list of
choices and can be re-selected if changed.
SET LINEAR
[Factory default = 0.01 VOLTS to 10 VOLTS corresponding to 1
mTORR to 1 TORR]
This will take the user to four different screens to setup the linear
analog output (see Analog Output section).
a) Set the minimum pressure
b) Set the minimum voltage corresponding to the minimum
pressure
c) Set the maximum pressure
d) Set the maximum voltage corresponding to the maximum
pressure
Note -
The LINEAR analog output provides a linear 0…10 V (dc)
output signal. The linear output voltage can be any value bet-
ween 10 mV and 10 V corresponding to displayed pressure
between 1 mTorr and 1000 Torr. However, the useful range of
the linear analog output is three decades. If your application
requires the analog output to cover a pressure range exceeding
three decades then consider using the non-linear or the log-
linear analog output. See ANALOG TYPE menu below to select
log-linear or non-linear analog output.
INFO
This screen shows the unit software version.